Closed-loop class E transcutaneous power and data link for microimplants
- PMID: 1601440
- DOI: 10.1109/10.141197
Closed-loop class E transcutaneous power and data link for microimplants
Abstract
Magnetic transcutaneous coupling is frequently used for power and data transfer to implanted electronic devices. The proposed development of MicroImplants, small enough to be injected through a hypodermic needle suggest the need for a high-efficiency magnetic transcutaneous link. This paper describes the use of a multifrequency transmitter coil driver based upon the Class E topology. The development of a "high-Q approximation" which simplifies the design procedure is presented. A closed-loop controller to compensate for transmitter and receiver variations, and a method of data modulation, using synchronous frequency shifting are described. The closed-loop Class E circuit shows great promise, especially for circuits with unusually low coefficients of coupling. Currents of several amperes, at radio frequencies, can easily and efficiently be obtained.
Similar articles
-
Class E driver for transcutaneous power and data link for implanted electronic devices.Med Biol Eng Comput. 1992 Jan;30(1):69-75. doi: 10.1007/BF02446196. Med Biol Eng Comput. 1992. PMID: 1640758
-
Micropower circuits for bidirectional wireless telemetry in neural recording applications.IEEE Trans Biomed Eng. 2005 Nov;52(11):1950-9. doi: 10.1109/TBME.2005.856247. IEEE Trans Biomed Eng. 2005. PMID: 16285399
-
Wireless technologies for closed-loop retinal prostheses.J Neural Eng. 2009 Dec;6(6):065004. doi: 10.1088/1741-2560/6/6/065004. Epub 2009 Oct 23. J Neural Eng. 2009. PMID: 19850974 Review.
-
Low-power transceiver analog front-end circuits for bidirectional high data rate wireless telemetry in medical endoscopy applications.IEEE Trans Biomed Eng. 2007 Jul;54(7):1291-9. doi: 10.1109/TBME.2006.889768. IEEE Trans Biomed Eng. 2007. PMID: 17605360
-
Automatic frequency controller for power amplifiers used in bio-implanted applications: issues and challenges.Sensors (Basel). 2014 Dec 11;14(12):23843-70. doi: 10.3390/s141223843. Sensors (Basel). 2014. PMID: 25615728 Free PMC article. Review.
Cited by
-
Simplified design equations for Class-E neural prosthesis transmitters.IEEE Trans Biomed Eng. 2013 May;60(5):1414-21. doi: 10.1109/TBME.2012.2237172. Epub 2013 Jan 1. IEEE Trans Biomed Eng. 2013. PMID: 23292784 Free PMC article.
-
Three-dimensional power receiver for in vivo robotic capsules.Med Biol Eng Comput. 2007 Oct;45(10):997-1002. doi: 10.1007/s11517-007-0223-9. Epub 2007 Aug 8. Med Biol Eng Comput. 2007. PMID: 17684783
-
Wearable and implantable wireless sensor network solutions for healthcare monitoring.Sensors (Basel). 2011;11(6):5561-95. doi: 10.3390/s110605561. Epub 2011 May 26. Sensors (Basel). 2011. PMID: 22163914 Free PMC article. Review.
-
A fully implantable rodent neural stimulator.J Neural Eng. 2012 Feb;9(1):014001. doi: 10.1088/1741-2560/9/1/014001. Epub 2012 Jan 17. J Neural Eng. 2012. PMID: 22248468 Free PMC article.
-
Robust wireless power transfer using a nonlinear parity-time-symmetric circuit.Nature. 2017 Jun 14;546(7658):387-390. doi: 10.1038/nature22404. Nature. 2017. PMID: 28617463
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources