Hemodynamic performance and incidence of patient-prosthesis mismatch of the complete supraannular perimount magna bioprosthesis in the aortic position
- PMID: 16037868
- DOI: 10.1055/s-2005-837678
Hemodynamic performance and incidence of patient-prosthesis mismatch of the complete supraannular perimount magna bioprosthesis in the aortic position
Abstract
Background: Complete supraannular placement and smaller stent design allow the implantation of a Perimount Magna bioprosthesis with a larger inner diameter than that of a standard Perimount. This study compares the hemodynamic performance and the incidence of patient-prosthesis mismatch (PPM) of both prostheses.
Methods: 128 patients underwent aortic valve replacement, receiving either a Magna (n = 57) or a standard (n = 71) prosthesis. Inner aortic annulus diameter was measured intraoperatively by a hegar dilator to match echocardiographically obtained results to the annulus diameter instead of matching them to labelled valve size.
Results: The Magna was significantly superior with respect to mean pressure gradient and effective orifice area in patients with an annulus diameter of 22-23 mm. In patients with an annulus diameter < 22 mm or > 23 mm, there was a non-significant trend towards superior hemodynamics in the Magna group. Severe PPM (effective orifice area index < or = 0.65 cm (2)/m (2)) was present in 11.1% (Magna) vs. 42.1% (Standard) of patients with an annulus diameter < 22 mm; in 0% (Magna) vs. 13.8% (Standard) with an annulus diameter of 22-23 mm; no PPM was seen in patients with annulus diameter > 23 mm in both groups.
Conclusions: The Perimount Magna had a significantly reduced incidence of patient-prosthesis mismatch and superior hemodynamics compared to the standard Perimount.
Similar articles
-
The effective orifice area/patient aortic annulus area ratio: a better way to compare different bioprostheses? A prospective randomized comparison of the Mosaic and Perimount bioprostheses in the aortic position.J Heart Valve Dis. 2004 May;13(3):382-8; discussion 388-9. J Heart Valve Dis. 2004. PMID: 15222284 Clinical Trial.
-
Hemodynamic comparison of bioprostheses for complete supra-annular position in patients with small aortic annulus.J Am Coll Cardiol. 2005 Jun 21;45(12):2054-60. doi: 10.1016/j.jacc.2005.03.039. J Am Coll Cardiol. 2005. PMID: 15963409 Clinical Trial.
-
Influence of completely supra-annular placement of bioprostheses on exercise hemodynamics in patients with a small aortic annulus.J Thorac Cardiovasc Surg. 2007 May;133(5):1234-41. doi: 10.1016/j.jtcvs.2006.10.074. J Thorac Cardiovasc Surg. 2007. PMID: 17467435
-
Patient-prosthesis mismatch: the Japanese experience.Ann Thorac Cardiovasc Surg. 2006 Jun;12(3):159-65. Ann Thorac Cardiovasc Surg. 2006. PMID: 16823327 Review.
-
Aortic root enlargement does not increase the surgical risk and short-term patient outcome?Eur J Cardiothorac Surg. 2011 Aug;40(2):441-7. doi: 10.1016/j.ejcts.2010.11.064. Epub 2011 Jan 13. Eur J Cardiothorac Surg. 2011. PMID: 21236693 Review.
Cited by
-
New design bioprosthesis: early outcome of Carpentier-Edwards PERIMOUNT Magna in the small annular aortic position.J Artif Organs. 2011 Dec;14(4):284-8. doi: 10.1007/s10047-011-0579-3. Epub 2011 Jun 21. J Artif Organs. 2011. PMID: 21692003
-
A mechanistic investigation of the EDWARDS INTUITY Elite valve's hemodynamic performance.Gen Thorac Cardiovasc Surg. 2020 Jan;68(1):9-17. doi: 10.1007/s11748-019-01154-y. Epub 2019 Jun 27. Gen Thorac Cardiovasc Surg. 2020. PMID: 31250203
-
Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method?Heart. 2007 May;93(5):615-20. doi: 10.1136/hrt.2006.102764. Epub 2006 Dec 12. Heart. 2007. PMID: 17164480 Free PMC article.
-
Aortic valve replacement with sutureless and rapid deployment aortic valve prostheses.J Geriatr Cardiol. 2016 Sep;13(6):504-10. doi: 10.11909/j.issn.1671-5411.2016.06.006. J Geriatr Cardiol. 2016. PMID: 27582765 Free PMC article.
-
Echocardiographic Assessment of Heart Valve Prostheses.J Cardiovasc Echogr. 2014 Oct-Dec;24(4):103-113. doi: 10.4103/2211-4122.147201. J Cardiovasc Echogr. 2014. PMID: 28465917 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources