A hydroxyapatite-coated total knee replacement: prospective analysis of 1000 patients
- PMID: 16049241
- DOI: 10.1302/0301-620X.87B8.15772
A hydroxyapatite-coated total knee replacement: prospective analysis of 1000 patients
Abstract
We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.
Comment in
-
A contemporary snapshot of the use of hydroxyapatite coating in orthopaedic surgery.J Bone Joint Surg Br. 2005 Aug;87(8):1046-9. doi: 10.1302/0301-620X.87B8.16692. J Bone Joint Surg Br. 2005. PMID: 16049236 Review. No abstract available.