Aortoiliac and lower extremity arteries assessed with 16-detector row CT angiography: prospective comparison with digital subtraction angiography

Affiliations
Free article
Comparative Study

Aortoiliac and lower extremity arteries assessed with 16-detector row CT angiography: prospective comparison with digital subtraction angiography

Jürgen K Willmann et al. Radiology. 2005 Sep.
Free article

Abstract

Purpose: To prospectively compare the accuracy of 16-detector row computed tomographic (CT) angiography with conventional digital subtraction angiography (DSA) as the reference standard in the assessment of aortoiliac and lower extremity arteries in patients with peripheral arterial disease.

Materials and methods: This study was approved by the institutional review board, and informed consent was obtained. A total of 39 consecutive patients (27 men [mean age, 66 years] and 12 women [mean age, 64 years]) with peripheral arterial disease underwent both conventional DSA and 16-detector row CT angiography. For data analysis, the arterial vascular system was divided into 35 segments. A total of 1365 arterial segments were analyzed for arterial stenosis by two independent blinded readers using a four-point grading system (grade 1, <10% luminal narrowing; grade 2, 10%-49% luminal narrowing; grade 3, 50%-99% luminal narrowing; grade 4, occlusion). Interobserver agreements were calculated by using kappa statistics. A third independent blinded reader assessed possible reasons for disagreements between 16-detector row CT angiographic findings and conventional DSA findings. Effective radiation dose was calculated for both imaging modalities.

Results: Sixteen-detector row CT angiographic and conventional DSA findings were diagnostic in all vascular segments. Compared with conventional DSA, the sensitivity and specificity of 16-detector row CT angiography with regard to detection of hemodynamically significant stenosis in all 35 arterial segments were 96% and 97%, respectively, for both readers. Readers 1 and 2 overestimated arterial stenosis in 42 (3%) and 34 (2%) arterial segments, respectively, and underestimated arterial stenosis in 13 (1%) and 10 (1%) arterial segments, respectively. Interobserver agreement was excellent (kappa = 0.84-1.00). Presence of anteroposteriorly located luminal narrowing and extensive vascular wall calcification were considered main reasons for disagreements between imaging modalities. Effective radiation dose was lower for 16-detector row CT angiography (1.6-3.9 mSv) than for conventional DSA (6.4-16.0 mSv).

Conclusion: Sixteen-detector row CT angiography is an accurate and reliable noninvasive alternative to conventional DSA in the assessment of aortoiliac and lower extremity arteries in patients with peripheral arterial disease.