False-positive alarms for bacterial screening of platelet concentrates with BacT/ALERT new-generation plastic bottles: a multicenter pilot study
- PMID: 16078911
- DOI: 10.1111/j.1537-2995.2005.00194.x
False-positive alarms for bacterial screening of platelet concentrates with BacT/ALERT new-generation plastic bottles: a multicenter pilot study
Abstract
Background: The microbial detection system BacT/ALERT (bioMérieux) is widely used to monitor bacterial contamination of platelet concentrates (PCs). Recently, the manufacturer introduced polycarbonate culture bottles and a modified pH-sensitive liquid emulsion sensor as microbial growth indicator. This reconfigured assay was investigated in a routine setting.
Study design and methods: In each of eight transfusion centers, samples from 500 consecutive PCs were monitored for 1 week. For all PCs with a positive BacT/ALERT signal, retained samples and, if available, original PC containers and concomitant red blood cell concentrates were analyzed independently. Initially BacT/ALERT-positive PCs without bacterial identification in any sample were defined as false-positive. BacT/ALERT-positive PCs with bacteria in the first sample only were called potentially positive. PCs with bacteria in the first sample and the same strain in at least one additional sample were accepted as positive.
Results: Five PCs (0.13%) were positive, 9 PCs (0.23%) were potentially positive, and 35 PCs (0.9%) were false-positive. The rate of false-positive BacT/ALERT results varied substantially between centers (<0.2%-3.2%). Tracings from false-positive cultures lacked an exponential increase of the signal during incubation. Most of these false-positives were due to malfunctioning cells in various BacT/ALERT incubation units.
Conclusion: Careful assessment of individual tracings of samples with positive signals helps to identify malfunctioning incubation units. Their early shutdown or replacement minimizes the high rate of unrectifiable product rejects attributed to false-positive alarms and avoids unnecessary concern of doctors and patients after conversion to a reconfigured BacT/ALERT assay.
Similar articles
-
Six years' experience of using the BacT/ALERT system to screen all platelet concentrates, and additional testing of outdated platelet concentrates to estimate the frequency of false-negative results.Vox Sang. 2005 Feb;88(2):93-7. doi: 10.1111/j.1423-0410.2005.00596.x. Vox Sang. 2005. PMID: 15720606
-
Bacterial contamination of platelet concentrates: results of a prospective multicenter study comparing pooled whole blood-derived platelets and apheresis platelets.Transfusion. 2007 Apr;47(4):644-52. doi: 10.1111/j.1537-2995.2007.01166.x. Transfusion. 2007. PMID: 17381623
-
Comparison of three bacterial detection methods under routine conditions.Vox Sang. 2007 Jan;92(1):15-21. doi: 10.1111/j.1423-0410.2006.00850.x. Vox Sang. 2007. PMID: 17181586
-
Applications of real-time PCR in the screening of platelet concentrates for bacterial contamination.Expert Rev Mol Diagn. 2006 Nov;6(6):865-72. doi: 10.1586/14737159.6.6.865. Expert Rev Mol Diagn. 2006. PMID: 17140373 Review.
-
Bacterial detection of platelets: current problems and possible resolutions.Transfus Med Rev. 2005 Oct;19(4):259-72. doi: 10.1016/j.tmrv.2005.05.002. Transfus Med Rev. 2005. PMID: 16214015 Review.
Cited by
-
Evaluation of Mirasol pathogen reduction system by artificially contaminating platelet concentrates with Staphylococcus epidermidis: A pilot study from India.Asian J Transfus Sci. 2016 Jul-Dec;10(2):127-31. doi: 10.4103/0973-6247.187946. Asian J Transfus Sci. 2016. PMID: 27605849 Free PMC article.
-
From the donor's arm to blood product: a study on bacterial contamination of apheresis platelet concentrates.Blood Transfus. 2007 Jul;5(3):130-3. doi: 10.2450/2007.0003-07. Blood Transfus. 2007. PMID: 19204765 Free PMC article.
-
Implementation of Bacterial Detection Methods into Blood Donor Screening - Overview of Different Technologies.Transfus Med Hemother. 2011;38(4):259-265. doi: 10.1159/000330305. Epub 2011 Jul 7. Transfus Med Hemother. 2011. PMID: 22016697 Free PMC article.
-
Characteristics of False-Positive Alarms in the BacT/Alert 3D System.Microbiol Spectr. 2022 Jun 29;10(3):e0005522. doi: 10.1128/spectrum.00055-22. Epub 2022 Apr 25. Microbiol Spectr. 2022. PMID: 35467361 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources