. 2005 Aug;43(8):3769-71.
doi: 10.1128/JCM.43.8.3769-3771.2005.

Evaluation of the VACUTAINER PPT Plasma Preparation Tube for use with the Bayer VERSANT assay for quantification of human immunodeficiency virus type 1 RNA

Affiliations

Evaluation of the VACUTAINER PPT Plasma Preparation Tube for use with the Bayer VERSANT assay for quantification of human immunodeficiency virus type 1 RNA

Tarek Elbeik et al. J Clin Microbiol. 2005 Aug.

Abstract

Separation and storage of plasma within 2 h of phlebotomy is required for the VACUTAINER PPT Plasma Preparation Tube (PPT) versus 4 h for the predecessor VACUTAINER EDTA tube for human immunodeficiency virus type 1 (HIV-1) viral load (HIVL) testing by the VERSANT HIV-1 RNA 3.0 assay (branched DNA). The 2-h limit for PPT imposes time constraints for handling and transporting to the testing laboratory. This study compares HIVL reproducibility from matched blood in EDTA tubes and PPTs and between PPT pairs following processing within 4 h of phlebotomy, stability of plasma HIV-1 RNA at 24- and 72-h room temperature storage in the tube, and comparative labor and supply requirements. Blood from 159 patients was collected in paired tubes (EDTA/PPT or PPT/PPT): 86 paired EDTA tubes and PPTs were processed 4 h following phlebotomy and their HIVLs were compared, 42 paired PPT/PPT pairs were analyzed for intertube HIVL reproducibility, and 31 PPT/PPT pairs were analyzed for HIV-1 RNA stability by HIVL. Labor and supply requirements were compared between PPT and EDTA tubes. PPTs produce results equivalent to standard EDTA tube results when processed 4 h after phlebotomy. PPT intertube analyte results are reproducible. An average decrease of 13% and 37% in HIVL was observed in PPT plasma after 24 and 72 h of room temperature storage, respectively; thus, plasma can be stored at room temperature up to 24 h in the original tube. PPTs offer labor and supply savings over EDTA tubes.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Comparison of HIVL values for blood collected in PPT versus EDTA tubes. One EDTA tube and one PPT were collected from each of 86 patients, and plasma was isolated and tested for HIVLs using the HIV-1 bDNA assay.
FIG. 2.
FIG. 2.
Comparison of HIVL values for blood collected in duplicate PPT. Two PPTs were collected from each of 42 patients, and plasma was isolated and tested for HIVLs using a HIV-1 bDNA assay.

Similar articles

Cited by

References

    1. Bland, J. M., and D. G. Altman. 1995. Comparing two methods of clinical measurement: a personal history. Int. J. Epidemiol. 24(Suppl. 1):S7-S14. - PubMed
    1. Collins, M. L., B. Irvine, D. Tyner, E. Fine, C. Zayati, C. Chang, T. Horn, D. Ahle, J. Detmer, L. P. Shen, J. Kolberg, S. Bushnell, M. S. Urdea, and D. D. Ho. 1997. A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml. Nucleic Acids Res. 25:2979-2984. - PMC - PubMed
    1. Gleaves, C. A., J. Welle, M. Campbell, T. Elbeik, V. Ng, P. E. Taylor, K. Kuramoto, S. Aceituno, E. Lewalski, B. Joppa, L. Sawyer, C. Schaper, D. McNairn, and T. Quinn. 2002. Multicenter evaluation of the Bayer VERSANT HIV-1 RNA 3.0 assay: analytical and clinical performance. J. Clin. Virol. 25:205-216. - PubMed
    1. Holodniy, M., L. Rainen, S. Herman, and B. Yen-Lieberman. 2000. Stability of plasma human immunodeficiency virus load in VACUTAINER PPT plasma preparation tubes during overnight shipment. J. Clin. Microbiol. 38:323-326. - PMC - PubMed
    1. Holodniy, M., L. Mole, B. Yen-Lieberman, D. Margolis, C. Starkey, R. Carroll, T. Spahlinger, J. Todd, and J. B. Jackson. 1995. Comparative stabilities of quantitative human immunodeficiency virus RNA in plasma from samples collected in VACUTAINER CPT, VACUTAINER PPT, and standard VACUTAINER tubes. J. Clin. Microbiol. 33:1562-1566. - PMC - PubMed

LinkOut - more resources