Respiration can be monitored by photoplethysmography with high sensitivity and specificity regardless of anaesthesia and ventilatory mode
- PMID: 16095458
- DOI: 10.1111/j.1399-6576.2005.00721.x
Respiration can be monitored by photoplethysmography with high sensitivity and specificity regardless of anaesthesia and ventilatory mode
Abstract
Background: Photoplethysmography (PPG) is a non-invasive optical technique used, for instance, in pulse oximetry. Beside the pulse synchronous component, PPG has a respiratory synchronous variation (PPGr). Efforts have been made to utilize this component for indirect monitoring of respiratory rate and volume. Assessment of the clinical usefulness as well as of the physiological background of PPGr is required. We evaluated if anaesthesia and positive-pressure ventilation would affect PPGr.
Methods: We recorded reflection mode PPGr, at the forearm, and the respiratory synchronous changes in central venous pressure (CVP), peripheral venous pressure (PVP) and arterial blood pressure (ABP) in 12 patients. Recordings for each patient were made on three occasions: awake with spontaneous breathing; anaesthetized with spontaneous breathing; and anaesthetized with positive-pressure ventilation. We analyzed the sensitivity, specificity, coherence and time relationship between the signals.
Results: PPGr sensitivity for breath detection was [mean (SD)] >86(21)% and specificity >96(12)%. Respiratory detection in the macrocirculation (CVP, PVP and ABP) showed a sensitivity >83(29)% and specificity >93(12)%. The coherence between signals was high (0.75-0.99). The three measurement situations did not significantly influence sensitivity, specificity or time shifts between the PPGr, PVP, ABP, and the reference CVP signal despite changes in physiological data between measurements.
Conclusion: A respiratory synchronous variation in PPG and all invasive pressure signals was detected. The reflection mode PPGr signal seemed to be a constant phenomenon related to respiration regardless of whether or not the subject was awake, anaesthetized or ventilated, which increases its clinical usefulness in respiratory monitoring.
Similar articles
-
Combined photoplethysmographic monitoring of respiration rate and pulse: a comparison between different measurement sites in spontaneously breathing subjects.Acta Anaesthesiol Scand. 2007 Oct;51(9):1250-7. doi: 10.1111/j.1399-6576.2007.01375.x. Epub 2007 Aug 15. Acta Anaesthesiol Scand. 2007. PMID: 17711563
-
Macrocirculation is not the sole determinant of respiratory induced variations in the reflection mode photoplethysmographic signal.Physiol Meas. 2003 Nov;24(4):925-37. doi: 10.1088/0967-3334/24/4/009. Physiol Meas. 2003. PMID: 14658783 Clinical Trial.
-
Respiratory variations in the photoplethysmographic waveform: acute hypovolaemia during spontaneous breathing is not detected.Physiol Meas. 2010 Jul;31(7):953-62. doi: 10.1088/0967-3334/31/7/006. Epub 2010 Jun 7. Physiol Meas. 2010. PMID: 20530847
-
Respiration signals from photoplethysmography.Anesth Analg. 2013 Oct;117(4):859-865. doi: 10.1213/ANE.0b013e31828098b2. Epub 2013 Feb 28. Anesth Analg. 2013. PMID: 23449854 Review.
-
Photoplethysmography and its application in clinical physiological measurement.Physiol Meas. 2007 Mar;28(3):R1-39. doi: 10.1088/0967-3334/28/3/R01. Epub 2007 Feb 20. Physiol Meas. 2007. PMID: 17322588 Review.
Cited by
-
Age and gender do not influence the ability to detect respiration by photoplethysmography.J Clin Monit Comput. 2006 Dec;20(6):431-6. doi: 10.1007/s10877-006-9050-z. Epub 2006 Oct 11. J Clin Monit Comput. 2006. PMID: 17033878
-
Toward a Robust Estimation of Respiratory Rate From Pulse Oximeters.IEEE Trans Biomed Eng. 2017 Aug;64(8):1914-1923. doi: 10.1109/TBME.2016.2613124. Epub 2016 Nov 18. IEEE Trans Biomed Eng. 2017. PMID: 27875128 Free PMC article.
-
Pulse oximetry-derived respiratory rate in general care floor patients.J Clin Monit Comput. 2015 Feb;29(1):113-20. doi: 10.1007/s10877-014-9575-5. Epub 2014 May 6. J Clin Monit Comput. 2015. PMID: 24796734 Free PMC article.
-
Measurement of respiratory rate from the photoplethysmogram in chest clinic patients.J Clin Monit Comput. 2007 Feb;21(1):55-61. doi: 10.1007/s10877-006-9059-3. Epub 2006 Nov 25. J Clin Monit Comput. 2007. PMID: 17131084
-
Identifying airway obstructions using photoplethysmography (PPG).J Clin Monit Comput. 2008 Apr;22(2):95-101. doi: 10.1007/s10877-008-9110-7. Epub 2008 Jan 25. J Clin Monit Comput. 2008. PMID: 18219579 Clinical Trial.