Abstract
The prospective Swedish Knee Arthroplasty Project contained 772 PCA, 1564 Marmor and 1441 St Georg unicompartment prostheses, implanted for gonarthrosis during the period December 1983-July 1990. Follow-up was performed in September 1991. There was a significant difference in the cumulative revision rate after only 2 years, increasing to 15 percent for the PCA, compared with 5-7 percent for the two other types. The difference between PCA and Marmor/St Georg was loosening of the femoral component. Major polyethylene wear was noted in a quarter of the revised PCA tibial components. The wear pattern of Marmor/St Georg was not analyzed in this study.
Similar articles
-
Polyethylene wear in unicondylar knee prostheses. 106 retrieved Marmor, PCA, and St Georg tibial components compared.Acta Orthop Scand. 1992 Jun;63(3):247-55. doi: 10.3109/17453679209154776. Acta Orthop Scand. 1992. PMID: 1609584
-
Oxford meniscal bearing knee versus the Marmor knee in unicompartmental arthroplasty for arthrosis. A Swedish multicenter survival study.J Arthroplasty. 1995 Dec;10(6):722-31. doi: 10.1016/s0883-5403(05)80066-x. J Arthroplasty. 1995. PMID: 8749752
-
[Uncemented knee prosthesis. Results apropos of 58 cases with a minimum of 5-year follow-up].Rev Chir Orthop Reparatrice Appar Mot. 1995;81(6):505-13. Rev Chir Orthop Reparatrice Appar Mot. 1995. PMID: 8560021 French.
-
Is an intact anterior cruciate ligament needed in order to have a well-functioning unicondylar knee replacement?Clin Orthop Relat Res. 2004 Nov;(428):170-3. doi: 10.1097/01.blo.0000148895.78766.8a. Clin Orthop Relat Res. 2004. PMID: 15534539 Review.
-
[Single-compartment "Lotus" type knee prosthesis in the treatment of lateralized gonarthrosis: results in 135 cases with a mean follow-up of 4.6 years].Rev Chir Orthop Reparatrice Appar Mot. 1993;79(7):565-76. Rev Chir Orthop Reparatrice Appar Mot. 1993. PMID: 8085038 Review. French.
Cited by
-
Influence of Preservation of Normal Knee Contact Stress on Other Compartments with respect to the Tibial Insert Design for Unicompartmental Knee Arthroplasty.Appl Bionics Biomech. 2019 Nov 14;2019:9246379. doi: 10.1155/2019/9246379. eCollection 2019. Appl Bionics Biomech. 2019. PMID: 31827605 Free PMC article.
-
Results of a French multicentre retrospective experience with four hundred and eighteen failed unicondylar knee arthroplasties.Int Orthop. 2013 Jul;37(7):1273-8. doi: 10.1007/s00264-013-1915-4. Epub 2013 May 29. Int Orthop. 2013. PMID: 23715952 Free PMC article.
-
Common causes of failed unicompartmental knee arthroplasty: a single-centre analysis of four hundred and seventy one cases.Int Orthop. 2014 May;38(5):961-5. doi: 10.1007/s00264-013-2263-0. Epub 2014 Jan 9. Int Orthop. 2014. PMID: 24402555 Free PMC article.
-
Does bearing design influence midterm survivorship of unicompartmental arthroplasty?Clin Orthop Relat Res. 2010 Jan;468(1):73-81. doi: 10.1007/s11999-009-0975-7. Epub 2009 Jul 14. Clin Orthop Relat Res. 2010. PMID: 19597898 Free PMC article.
-
The Swedish Knee Arthroplasty Register: a review.Bone Joint Res. 2014 Jul;3(7):217-22. doi: 10.1302/2046-3758.37.2000289. Bone Joint Res. 2014. PMID: 24986492 Free PMC article. Review.