The use of a head-mounted display for visualization in neuroendoscopy
Abstract
Objective: To evaluate the clinical use of a head-mounted display (HMD) for visualization in all neuroendoscopic procedures.
Materials and methods: We retrospectively evaluated all endoscopic and endoscope-controlled procedures in which the HMD was used in our department between July 1999 and June 2002.
Results: A total of 269 endoscopic procedures were performed. In 147 cases intraventricular endoscopic procedures were carried out, mostly third ventriculocisternostomies, for which a fiberscope was used exclusively. Thirty intracranial cysts were fenestrated or removed (colloid cysts) with the help of various endoscopes. A total of 87 endoscopic transsphenoidal surgeries were performed with a lensscope. In only one case was it necessary to abandon use of the HMD due to inferior visualization; in all other cases visualization by the HMD was thought to be sufficient. Fatigue of the surgeon due to wearing the helmet did not occur. All surgeons had the impression that visual strain was decreased in comparison to looking at a monitor from a distance. The working position was considered to be more comfortable when wearing the HMD, and eye-hand coordination was improved. No technical problems occurred with the system.
Conclusions: The HMD is a new visualization tool in neurosurgery that may improve the ergonomics of neuroendoscopic and endoscope-controlled procedures.
Similar articles
-
[Clinical application of minimally invasive neuroendoscopic techniques].Zhongguo Yi Xue Ke Xue Yuan Xue Bao. 2005 Feb;27(1):22-5. Zhongguo Yi Xue Ke Xue Yuan Xue Bao. 2005. PMID: 15782487 Chinese.
-
Multiple brain tumor nodule resections under direct visualization of a neuronavigated endoscope.Minim Invasive Neurosurg. 2007 Aug;50(4):227-32. doi: 10.1055/s-2007-985861. Minim Invasive Neurosurg. 2007. PMID: 17948182
-
Three-dimensional neurostereoendoscopy: subjective and objective comparison to 2D.Minim Invasive Neurosurg. 2009 Feb;52(1):25-31. doi: 10.1055/s-0028-1104567. Epub 2009 Feb 26. Minim Invasive Neurosurg. 2009. PMID: 19247901
-
Current state and future development of intracranial neuroendoscopic surgery.Expert Rev Med Devices. 2005 May;2(3):351-73. doi: 10.1586/17434440.2.3.351. Expert Rev Med Devices. 2005. PMID: 16288598 Review.
-
General principles and intraventricular neuroendoscopy: endoscopic techniques.World Neurosurg. 2013 Feb;79(2 Suppl):S14.e23-8. doi: 10.1016/j.wneu.2012.02.031. Epub 2012 Feb 10. World Neurosurg. 2013. PMID: 22381832 Review.
Cited by
-
Three-dimensional visualisation and articulating instrumentation: Impact on simulated laparoscopic tasks.J Minim Access Surg. 2008 Apr;4(2):31-8. doi: 10.4103/0972-9941.41938. J Minim Access Surg. 2008. PMID: 19547678 Free PMC article.
-
See-Through Type 3D Head-Mounted Display-Based Surgical Microscope System for Microsurgery: A Feasibility Study.JMIR Mhealth Uhealth. 2019 Mar 7;7(3):e11251. doi: 10.2196/11251. JMIR Mhealth Uhealth. 2019. PMID: 30843867 Free PMC article.
-
Can a Wireless Full-HD Head Mounted Display System Improve Knee Arthroscopy Performance? - A Randomized Study Using a Knee Simulator.Surg Innov. 2023 Aug;30(4):477-485. doi: 10.1177/15533506221142960. Epub 2022 Nov 30. Surg Innov. 2023. PMID: 36448618 Free PMC article. Clinical Trial.
-
Compact Smartphone-Based Laser Speckle Contrast Imaging Endoscope Device for Point-of-Care Blood Flow Monitoring.Biosensors (Basel). 2022 Jun 9;12(6):398. doi: 10.3390/bios12060398. Biosensors (Basel). 2022. PMID: 35735546 Free PMC article.
-
Feasibility of Wearable Display Glasses for Medical Students in the Endoscopy Room.Clin Endosc. 2021 Sep;54(5):694-700. doi: 10.5946/ce.2020.246. Epub 2021 Mar 15. Clin Endosc. 2021. PMID: 33715343 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources