Different conditions that could result in the bispectral index indicating an incorrect hypnotic state
- PMID: 16115989
- DOI: 10.1213/01.ane.0000167269.62966.af
Different conditions that could result in the bispectral index indicating an incorrect hypnotic state
Abstract
Since its introduction in 1996, the Bispectral Index (BIS) has gained increasing popularity in daily anesthesia practice. However, numerous reports have been appearing in the literature of paradoxical BIS changes and inaccurate readings. The purpose of this review is to assess the utility of BIS monitoring through examining the various published reports of all BIS values not coinciding with a clinically judged sedative-hypnotic state, whether arising from an underlying pathophysiology of electroencephalographic (EEG) cerebral function or because of shortcomings in the performance and design of the BIS monitor. High electromyographic activity and electric device interference could create subtle artifact signal pollution without their necessarily being displayed as artifacts. This would be misinterpreted by the BIS algorithm as EEG activity and assigned a spuriously increased BIS value. Numerous clinical conditions that have a direct effect on EEG cerebral function could also directly influence the BIS value.
Comment in
- 
  
  Cardiopulmonary bypass machine can interfere with accuracy of BIS monitor.Anesth Analg. 2007 Aug;105(2):534; author reply 534-5. doi: 10.1213/01.ane.0000265661.56155.08. Anesth Analg. 2007. PMID: 17646520 No abstract available.
References
- 
    - Guingo LD, Chabot RJ, Prichep LS, et al. Quantitative EEG changes associated with loss and return of consciousness in healthy adult volunteers anaesthetized with propofol or sevoflorane. Br J Anaesth 2001;87:421–8.
 
- 
    - Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology 1998;89:980–1002.
 
- 
    - Rampil IJ, Kim J, Lenhardt R, et al. Bispectral EEG index during nitrous oxide administration. Anesthesiology 1998;89:671–7.
 
- 
    - Puri GD. Paradoxical changes in bispectral index during nitrous oxide administration. Br J Anaesth 2001;86:141–2.
 
- 
    - Henrie JR, Parkhouse J, Bickford RG. Alteration of human consciousness by nitrous oxide as assessed by electroencephalography and psychological tests. Anesthesiology 1961;22:247–59.
 
Publication types
MeSH terms
Substances
LinkOut - more resources
- Full Text Sources
- Medical
- Miscellaneous
 
        