Clinical Trial
. 2005 Sep;101(3):910-915.
doi: 10.1213/01.ane.0000166975.38649.27.

Cervical spine motion: a fluoroscopic comparison during intubation with lighted stylet, GlideScope, and Macintosh laryngoscope

Affiliations
Clinical Trial

Cervical spine motion: a fluoroscopic comparison during intubation with lighted stylet, GlideScope, and Macintosh laryngoscope

Timothy P Turkstra et al. Anesth Analg. 2005 Sep.

Erratum in

  • Anesth Analg. 2005 Oct;101(4):1011

Abstract

The question of which is the optimum technique to intubate the trachea in a patient who may have a cervical(C)-spine injury remains unresolved. We compared, using fluoroscopic video, C-spine motion during intubation for Macintosh 3 blade, GlideScope, and Intubating Lighted Stylet, popularly known as the Lightwand or Trachlight. Thirty-six healthy patients were randomized to participate in a crossover trial of either Lightwand or GlideScope to Macintosh laryngoscopy, with in-line stabilization. C-spine motion was examined at the Occiput-C1 junction, C1-2 junction, C2-5 motion segment, and C5-thoracic motion segment during manual ventilation via bag-mask, laryngoscopy, and intubation. Time to intubate was also measured. C-spine motion during bag-mask ventilation was 82% less at the four motion segments studied than during Macintosh laryngoscopy (P < 0.001). C-spine motion using the Lightwand was less than during Macintosh laryngoscopy, averaging 57% less at the four motion segments studied (P < 0.03). There was no significant difference in time to intubate between the Lightwand and the Macintosh blade. C-spine motion was reduced 50% at the C2-5 segment using the GlideScope (P < 0.04) but unchanged at the other segments. Laryngoscopy with GlideScope took 62% longer than with the Macintosh blade (P < 0.01). Thus, the Lightwand (Intubating Lighted Stylet) is associated with reduced C-spine movement during endotracheal intubation compared with the Macintosh laryngoscope.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Crosby E. Airway management after upper cervical spine injury: what have we learned? Can J Anaesth 2002;49:733–44.
    1. Brimacombe J, Keller C, Kunzel KH, et al. Cervical spine motion during airway management: a cinefluoroscopic study of posteriorly destabilized third cervical vertebrae in human cadavers. Anesth Analg 2000;91:1274–8.
    1. Walt B, Melischek M, Kabon B, et al. Tracheal intubation and cervical spine excursion: direct laryngoscopy vs. intubating laryngeal mask. Anaesthesia 2001;56:221–6.
    1. Hauswald M, Sklar DP, Tandberg D, Garcia JF. Cervical spine movement during airway management: cinefluoroscopic appraisal in human cadavers. Am J Emerg Med 1991;9:535–8.
    1. Lennarson PJ, Smith DW, Sawin PD, et al. Cervical spinal motion during intubation: efficacy of stabilization maneuvers in the setting of complete segmental instability. J Neurosurg 2001;94:265–70.

Publication types

LinkOut - more resources