Anodic plasma chemical treatment of titanium Schanz screws reduces pin loosening
- PMID: 16118562
- DOI: 10.1097/01.bot.0000164592.96724.a9
Anodic plasma chemical treatment of titanium Schanz screws reduces pin loosening
Abstract
Objective: This study was designed to assess the benefits of a new Anodic Plasma Chemical calcium-phosphate (APC-CaP) surface treatment on reducing pin track infection and pin loosening in comparison to anodized titanium (Ti) during external fracture fixation.
Methods: A tibial midshaft, transverse, 6-mm gap osteotomy was created in 17 adult female Swiss alpine sheep. The tibia was stabilized with an external fixator and 4 Schanz screws of Ti or APC-CaP-treated Ti. The sheep were examined during a 12-week observation period. Infection was assessed with weekly clinical pin track grading and microbiologic assessment at sacrifice. Pin loosening was assessed by grading for radiolucency on biweekly radiographs and by measuring extraction torque on pin removal. In vivo bending stiffness measurements were performed to determine gap healing. A qualitative histologic assessment of the tissue adjacent to pin sites was also performed.
Results: A trend (P = 0.056) for less infection around APC-CaP pins was found at 6 weeks, but the strength of this difference diminished with time. Significantly more radiolucency was found around Ti pins after 8 (P = 0.011) and 12 (P < 0.001) weeks. At all pin sites, the extraction torque for APC-CaP pins was higher than for Ti pins (P = 0.007). No difference in the progression of gap healing was found. Histology showed bone growth at the implant surface in the form of distance osteogenesis for Ti and contact osteogenesis for APC-CaP.
Conclusions: This study has shown that the APC-CaP surface improves the clinical performance of Ti pins with respect to pin loosening and pin track infection.
Similar articles
-
An in Vivo Experimental Comparison of Stainless Steel and Titanium Schanz Screws for External Fixation.Eur J Trauma Emerg Surg. 2007 Feb;33(1):59-68. doi: 10.1007/s00068-007-6053-5. Epub 2007 Feb 27. Eur J Trauma Emerg Surg. 2007. PMID: 26815976
-
Hydroxyapatite-coated Schanz pins in external fixators used for distraction osteogenesis : a randomized, controlled trial.J Bone Joint Surg Am. 2002 Jul;84(7):1162-6. doi: 10.2106/00004623-200207000-00011. J Bone Joint Surg Am. 2002. PMID: 12107316 Clinical Trial.
-
Increased insertion torque delays pin-bone interface loosening in external fixation with tapered bone screws.J Orthop Trauma. 2004 Oct;18(9):617-22. doi: 10.1097/00005131-200410000-00007. J Orthop Trauma. 2004. PMID: 15448451
-
State of the art review: techniques to avoid pin loosening and infection in external fixation.J Orthop Trauma. 2002 Mar;16(3):189-95. doi: 10.1097/00005131-200203000-00009. J Orthop Trauma. 2002. PMID: 11880783 Review.
-
Hydroxyapatite-coated external fixation pins.Expert Rev Med Devices. 2005 Jul;2(4):465-71. doi: 10.1586/17434440.2.4.465. Expert Rev Med Devices. 2005. PMID: 16293085 Review.
Cited by
-
Effects of Ti2448 half-pin with low elastic modulus on pin loosening in unilateral external fixation.J Mater Sci Mater Med. 2011 Jun;22(6):1579-88. doi: 10.1007/s10856-011-4313-8. Epub 2011 Apr 13. J Mater Sci Mater Med. 2011. PMID: 21487787
-
Effect of micro-arc oxidation surface modification of 3D-printed porous titanium alloys on biological properties.Ann Transl Med. 2022 Jun;10(12):710. doi: 10.21037/atm-22-2536. Ann Transl Med. 2022. PMID: 35845487 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous