Motion detection and correction for dynamic ( 15)O-water myocardial perfusion PET studies
- PMID: 16142471
- DOI: 10.1007/s00259-005-1846-4
Motion detection and correction for dynamic ( 15)O-water myocardial perfusion PET studies
Abstract
Purpose: Patient motion during dynamic PET studies is a well-documented source of errors. The purpose of this study was to investigate the incidence of frame-to-frame motion in dynamic ( 15)O-water myocardial perfusion PET studies, to test the efficacy of motion correction methods and to study whether implementation of motion correction would have an impact on the perfusion results.
Methods: We developed a motion detection procedure using external radioactive skin markers and frame-to-frame alignment. To evaluate motion, marker coordinates inside the field of view were determined in each frame for each study. The highest number of frames with identical spatial coordinates during the study were defined as "non-moved". Movement was considered present if even one marker changed position, by one pixel/frame compared with reference, in one axis, and such frames were defined as "moved". We tested manual, in-house-developed motion correction software and an automatic motion correction using a rigid body point model implemented in MIPAV (Medical Image Processing, Analysis and Visualisation) software. After motion correction, remaining motion was re-analysed. Myocardial blood flow (MBF) values were calculated for both non-corrected and motion-corrected datasets.
Results: At rest, patient motion was found in 18% of the frames, but during pharmacological stress the fraction increased to 45% and during physical exercise it rose to 80%. Both motion correction algorithms significantly decreased (p<0.006) the number of moved frames and the amplitude of motion (p<0.04). Motion correction significantly increased MBF results during bicycle exercise (p<0.02). At rest or during adenosine infusion, the motion correction had no significant effects on MBF values.
Conclusion: Significant motion is a common phenomenon in dynamic cardiac studies during adenosine infusion but especially during exercise. Applying motion correction for the data acquired during exercise clearly changed the MBF results, indicating that motion correction is required for these studies.
Similar articles
-
Motion Correction and Its Impact on Absolute Myocardial Blood Flow Measures with PET.Curr Cardiol Rep. 2018 Mar 24;20(5):34. doi: 10.1007/s11886-018-0977-8. Curr Cardiol Rep. 2018. PMID: 29574494 Review.
-
Effects of patient movement on measurements of myocardial blood flow and viability in resting ¹⁵O-water PET studies.J Nucl Cardiol. 2012 Jun;19(3):524-33. doi: 10.1007/s12350-012-9522-0. J Nucl Cardiol. 2012. PMID: 22314554 Free PMC article.
-
Improvement of algorithm for quantification of regional myocardial blood flow using 15O-water with PET.J Nucl Med. 2004 Nov;45(11):1908-16. J Nucl Med. 2004. PMID: 15534062 Clinical Trial.
-
Generation of parametric image of regional myocardial blood flow using H(2)(15)O dynamic PET and a linear least-squares method.J Nucl Med. 2005 Oct;46(10):1687-95. J Nucl Med. 2005. PMID: 16204719
-
Cardiac positron emission tomography imaging.Semin Nucl Med. 2005 Jan;35(1):17-36. doi: 10.1053/j.semnuclmed.2004.09.002. Semin Nucl Med. 2005. PMID: 15645392 Review.
Cited by
-
Motion Correction and Its Impact on Absolute Myocardial Blood Flow Measures with PET.Curr Cardiol Rep. 2018 Mar 24;20(5):34. doi: 10.1007/s11886-018-0977-8. Curr Cardiol Rep. 2018. PMID: 29574494 Review.
-
Cardiac 15O-water PET: Does mismatched attenuation correction not matter?J Nucl Cardiol. 2022 Jun;29(3):1129-1131. doi: 10.1007/s12350-021-02573-2. Epub 2021 Mar 9. J Nucl Cardiol. 2022. PMID: 33751477 No abstract available.
-
Detection and correction of patient motion in dynamic 15O-water PET MPI.J Nucl Cardiol. 2023 Dec;30(6):2736-2749. doi: 10.1007/s12350-023-03358-5. Epub 2023 Aug 28. J Nucl Cardiol. 2023. PMID: 37639181 Free PMC article.
-
Alignment of 3-dimensional cardiac structures in O-15-labeled water PET emission images with mutual information.J Nucl Cardiol. 2007 Jan;14(1):82-91. doi: 10.1016/j.nuclcard.2006.10.020. J Nucl Cardiol. 2007. PMID: 17276310
-
Effects of patient movement on measurements of myocardial blood flow and viability in resting ¹⁵O-water PET studies.J Nucl Cardiol. 2012 Jun;19(3):524-33. doi: 10.1007/s12350-012-9522-0. J Nucl Cardiol. 2012. PMID: 22314554 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources