Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2006 Sep;20(9):1030-3.
doi: 10.1038/sj.eye.6702073. Epub 2005 Sep 9.

Assessment of the reproducibility and repeatability of a method of grading macular subretinal fluid using optical coherence tomography

Affiliations

Assessment of the reproducibility and repeatability of a method of grading macular subretinal fluid using optical coherence tomography

S E Benson et al. Eye (Lond). 2006 Sep.

Abstract

Aim: To devise a quantative method for the measurement of the extent of macular subretinal fluid using optical coherence tomography (OCT), and to assess the interobserver and intraobserver agreement for this grading system.

Methods: Observational cohort series. Patients were a cohort who underwent retinal detachment surgery over an 18-month period. All patients had OCT scan at 6 weeks after surgery. The scans were graded by two independent observers experienced in OCT interpretation and each grader was masked to the others findings. Observer 1 then regraded the scans on a later day masked to his previous findings. The interobserver and intraobserver agreement was assessed using weighted Kappa (Kw) statistics.

Results: In all, 116 patients were analysed. Both the intraobserver and interobserver agreement was very high, with Kw being 0.9631 and 0.9070, respectively.

Conclusions: The grading system for assessment of the extent of macular subretinal fluid using OCT appears to have very good reproducibility and repeatability. We propose that this grading system would be clinically useful when applied to pathologies visible on OCT scan of the macula.