Mid- and long-term device migration after endovascular abdominal aortic aneurysm repair: a comparison of AneuRx and Zenith endografts
- PMID: 16171578
- DOI: 10.1016/j.jvs.2005.05.040
Mid- and long-term device migration after endovascular abdominal aortic aneurysm repair: a comparison of AneuRx and Zenith endografts
Abstract
Background: Freedom from migration is key to the durability of endovascular aneurysm repair (EVAR). This study evaluates the mid- and long-term incidence of migration with two different endografts.
Methods: Between September 1997 and June 2004, 235 patients were scheduled for EVAR with an AneuRx (Medtronic/AVE Inc.) or Zenith (Cook) endograft. Patients with fusiform, infrarenal aneurysms and a minimum 12 months of follow-up were analyzed, for a final cohort of 130 patients. Migration was assessed on axial computed tomography (CT) (2.5 to 3 mm cuts) as the distance from the most caudal renal artery to the first slice containing endograft (AneuRx) or to the top of the bare suprarenal stent (Zenith). Aortic neck diameters were measured at the most caudal renal artery. The initial postoperative CT scan was the baseline. Migration was defined by caudal movement of the endograft at two thresholds, > or =5 mm and > or =10 mm, or any migration with a related clinical event.
Results: Life-table analysis demonstrated AneuRx freedom from migration (> or =10 mm or clinical event) was 96.1%, 89.5%, 78.0%, and 72.0% at 1, 2, 3, and 4 years, respectively. Zenith freedom from migration was 100%, 97.6%, 97.6%, and 97.6% at 1, 2, 3, and 4 years, respectively (P = .01, log-rank test). The stricter 5-mm migration threshold found 67.4% of AneuRx and 90.1% of Zenith patients free from migration at 4 years of follow-up. Twelve out of 14 (85.7%) AneuRx patients (12/14) with migration (> or =10 mm or clinical event) underwent 14 related secondary procedures (13 endovascular, 1 open conversion). The single Zenith patient with migration (> or =10 mm) has not required adjuvant treatment. Mean follow-up was 39.0 +/- 2.3 months (AneuRx) and 30.8 +/- 1.9 months (Zenith, P = .01). Patients with and without migration did not differ in age, gender ratio, aneurysm diameter, and neck diameter. However, initial neck length was shorter in patients with migration (22.1 +/- 2.1 mm vs 31.2 +/- 1.2 mm, P = .02). A subset of patients (21.6%) experienced significant (defined as > or =3 mm) maximum aortic neck dilation. Of the AneuRx patients, > or =3 mm aortic neck dilation affected 30.8% of migrators vs 13.0% of nonmigrators (P = .20).
Conclusions: Endograft migration is a time-dependent phenomenon affected by both device choice and aortic neck length. A great majority of patients (85.7%) with migration of the AneuRx device ultimately required treatment. A minority of patients experienced aortic neck dilation that could be considered clinically significant. Careful surveillance for migration is an essential component of long-term follow-up after EVAR.
Similar articles
-
AneuRx device migration: incidence, risk factors, and consequences.Ann Vasc Surg. 2005 Mar;19(2):178-85. doi: 10.1007/s10016-004-0166-7. Ann Vasc Surg. 2005. PMID: 15782271
-
Long-term influence of suprarenal or infrarenal fixation on proximal neck dilatation and stentgraft migration after EVAR.Ann Vasc Surg. 2011 Nov;25(8):1012-9. doi: 10.1016/j.avsg.2010.08.013. Ann Vasc Surg. 2011. PMID: 22023937
-
Endograft migration one to four years after endovascular abdominal aortic aneurysm repair with the AneuRx device: a cautionary note.J Vasc Surg. 2002 Sep;36(3):476-84. doi: 10.1067/mva.2002.126561. J Vasc Surg. 2002. PMID: 12218970
-
Zenith abdominal aortic aneurysm endovascular graft: a literature review.J Cardiovasc Surg (Torino). 2009 Apr;50(2):165-70. J Cardiovasc Surg (Torino). 2009. PMID: 19329913 Review.
-
Endograft accommodation on the aortic bifurcation: an overview of anatomical fixation and implications for long-term stent-graft stability.J Endovasc Ther. 2011 Aug;18(4):462-70. doi: 10.1583/11-3411.1. J Endovasc Ther. 2011. PMID: 21861731 Review.
Cited by
-
Complementary roles of open and hybrid approaches to thoracoabdominal aortic aneurysm repair.J Vasc Surg. 2016 Nov;64(5):1228-1238. doi: 10.1016/j.jvs.2016.04.022. Epub 2016 Jul 18. J Vasc Surg. 2016. PMID: 27444368 Free PMC article.
-
Peer reviewed publications in 2005.Ochsner J. 2006 Spring;6(1):43-5. Ochsner J. 2006. PMID: 21765791 Free PMC article. No abstract available.
-
A Single-Institution Case Series of Total Endovascular Relining for Type 3 Endoleaks in Traditional Endovascular Aneurysm Repair (EVAR) Grafts with Raised Bifurcations.Ann Vasc Surg. 2024 Feb;99:332-340. doi: 10.1016/j.avsg.2023.08.038. Epub 2023 Oct 13. Ann Vasc Surg. 2024. PMID: 37839654 Free PMC article.
-
Aortic endograft infection after thoracic endovascular aortic repair: two case reports and literature review.Front Cardiovasc Med. 2025 Feb 25;12:1549613. doi: 10.3389/fcvm.2025.1549613. eCollection 2025. Front Cardiovasc Med. 2025. PMID: 40071233 Free PMC article.
-
2022 ACC/AHA Guideline for the Diagnosis and Management of Aortic Disease: A Report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines.J Am Coll Cardiol. 2022 Dec 13;80(24):e223-e393. doi: 10.1016/j.jacc.2022.08.004. Epub 2022 Nov 2. J Am Coll Cardiol. 2022. PMID: 36334952 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical