Inappropriate detection and therapy from an implantable cardioverter defibrillator due to atrial undersensing
- PMID: 16176546
- DOI: 10.1111/j.1540-8159.2005.00207.x
Inappropriate detection and therapy from an implantable cardioverter defibrillator due to atrial undersensing
Abstract
We present a case of inappropriate detection and therapy in a patient with an implantable cardioverter defibrillator. The device interpreted a supraventricular tachycardia as ventricular tachycardia due to unsensed P waves as a result of a relatively high level of atrial sensitivity setting. Subsequently, inappropriate therapy was delivered. The problem was resolved by decreasing the atrial sensitivity value.