Morphometric assessment of normal, suspect and glaucomatous optic discs with Stratus OCT and HRT II
- PMID: 16179933
- DOI: 10.1038/sj.eye.6702101
Morphometric assessment of normal, suspect and glaucomatous optic discs with Stratus OCT and HRT II
Abstract
Aims: To compare morphometric parameters and diagnostic performance of the new Stratus Optical Coherence Tomograph (OCT) Disc mode and the Heidelberg Retina Tomograph (HRT); to evaluate OCT's accuracy in determining optic nerve head (ONH) borders.
Methods: Controls and patients with ocular hypertension, glaucoma-like discs, and glaucoma were imaged with OCT Disc mode, HRT II, and colour disc photography (DISC-PHOT). In a separate session, automatically depicted ONH shape and size in OCT were compared with DISC-PHOT, and disc borders adjusted manually where required. In a masked fashion, all print-outs and photographs were studied and discs classified as normal, borderline, and abnormal. The Cohen kappa method was then applied to test for agreement of classification. Bland-Altman analysis was used for comparison of disc measures.
Results: In all, 49 eyes were evaluated. Automated disc margin recognition failed in 53%. Misplaced margin points were more frequently found in myopic eyes, but only 31/187 were located in an area of peripapillary atrophy. Agreement of OCT with photography-based diagnosis was excellent in normally looking ONHs, but moderate in discs with large cups, where HRT performed better. OCT values were consistently larger than HRT values for disc and cup area. Compared with HRT, small rim areas and volumes tended to be minimized by OCT, and larger ones to be magnified.
Conclusions: Stratus OCT Disc protocol performed overall well in differentiating between normal and glaucomatous ONHs. However, failure of disc border recognition was frequently observed, making manual correction necessary. ONH measures cannot be directly compared between HRT and OCT.
Similar articles
-
Disagreement between Heidelberg Retina Tomograph and optical coherence tomography in assessing optic nerve head configuration of eyes with band atrophy and normal eyes.Br J Ophthalmol. 2008 Oct;92(10):1382-6. doi: 10.1136/bjo.2007.134098. Epub 2008 Jul 28. Br J Ophthalmol. 2008. PMID: 18662917
-
Evaluation of optical coherence tomography and heidelberg retinal tomography parameters in detecting early and moderate glaucoma.Invest Ophthalmol Vis Sci. 2007 Jul;48(7):3138-45. doi: 10.1167/iovs.06-1407. Invest Ophthalmol Vis Sci. 2007. PMID: 17591883
-
Comparison of retinal nerve fiber layer thickness values using Stratus Optical Coherence Tomography and Heidelberg Retina Tomograph-III.J Glaucoma. 2009 Sep;18(7):528-34. doi: 10.1097/IJG.0b013e318193c29f. J Glaucoma. 2009. PMID: 19745667
-
[Imagings of glaucoma: characteristics of ocular configurations and automated diagnosis].Nippon Ganka Gakkai Zasshi. 2011 Mar;115(3):276-95; discussion 296. Nippon Ganka Gakkai Zasshi. 2011. PMID: 21476311 Review. Japanese.
-
[Modern diagnostic methods for suspected glaucoma and glaucoma].Klin Monbl Augenheilkd. 2004 Apr;221(4):227-46. doi: 10.1055/s-2004-812941. Klin Monbl Augenheilkd. 2004. PMID: 15118952 Review. German.
Cited by
-
Ability of cirrus HD-OCT optic nerve head parameters to discriminate normal from glaucomatous eyes.Ophthalmology. 2011 Feb;118(2):241-8.e1. doi: 10.1016/j.ophtha.2010.06.036. Epub 2010 Oct 28. Ophthalmology. 2011. PMID: 20920824 Free PMC article.
-
Spectral-Domain Optical Coherence Tomography for Glaucoma Diagnosis.Open Ophthalmol J. 2015 May 15;9:68-77. doi: 10.2174/1874364101509010068. eCollection 2015. Open Ophthalmol J. 2015. PMID: 26069519 Free PMC article.
-
Assessment of the optic disc morphology using spectral-domain optical coherence tomography and scanning laser ophthalmoscopy.Biomed Res Int. 2014;2014:275654. doi: 10.1155/2014/275654. Epub 2014 Jul 6. Biomed Res Int. 2014. PMID: 25110668 Free PMC article.
-
Agreement between optical coherence tomography and digital stereophotography in vertical cup-to-disc ratio measurement.Graefes Arch Clin Exp Ophthalmol. 2009 Mar;247(3):377-83. doi: 10.1007/s00417-008-0968-3. Epub 2008 Oct 29. Graefes Arch Clin Exp Ophthalmol. 2009. PMID: 18958489
-
Adaptive optics optical coherence tomography in glaucoma.Prog Retin Eye Res. 2017 Mar;57:76-88. doi: 10.1016/j.preteyeres.2016.11.001. Epub 2016 Dec 1. Prog Retin Eye Res. 2017. PMID: 27916682 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical