[Verification of normative values for the Swiss version of the Freiburg speech intelligibility test]
- PMID: 16189713
- DOI: 10.1007/s00106-005-1337-8
[Verification of normative values for the Swiss version of the Freiburg speech intelligibility test]
Abstract
Background and objective: In the Swiss version of the Freiburg speech intelligibility test five test words from the original German recording which are rarely used in Switzerland have been exchanged. Furthermore, differences in the transfer functions between headphone and loudspeaker presentation are not taken into account during calibration. New settings for the levels of the individual test words in the recommended recording and small changes in calibration procedures led us to make a verification of the currently used normative values.
Patients and methods: Speech intelligibility was measured in 20 subjects with normal hearing using monosyllabic words and numbers via headphones and loudspeakers.
Results: On average, 50% speech intelligibility was reached at levels which were 7.5 dB lower under free-field conditions than for headphone presentation. The average difference between numbers and monosyllabic words was found to be 9.6 dB, which is considerably lower than the 14 dB of the current normative curves.
Conclusions: There is a good agreement between our measurements and the normative values for tests using monosyllabic words and headphones, but not for numbers or free-field measurements.
Similar articles
-
[The Freiburg speech intelligibility test : A pillar of speech audiometry in German-speaking countries].HNO. 2016 Aug;64(8):540-8. doi: 10.1007/s00106-016-0150-x. HNO. 2016. PMID: 27259640 Review. German.
-
[Evaluation of the perceptual equivalence of test lists in the Freiburg monosyllabic speech test].HNO. 2016 Aug;64(8):572-83. doi: 10.1007/s00106-016-0192-0. HNO. 2016. PMID: 27418349 German.
-
[Test-retest reliability of the Freiburg monosyllabic speech test].HNO. 2016 Aug;64(8):564-71. doi: 10.1007/s00106-016-0166-2. HNO. 2016. PMID: 27286728 German.
-
[The Freiburg monosyllable word test in postoperative cochlear implant diagnostics].HNO. 2016 Aug;64(8):601-7. doi: 10.1007/s00106-016-0194-y. HNO. 2016. PMID: 27393292 German.
-
[Use of Freiburg monosyllabic test words in the contemporary German language : Currentness of the test words].HNO. 2016 Aug;64(8):549-56. doi: 10.1007/s00106-016-0163-5. HNO. 2016. PMID: 27286726 Review. German.
Cited by
-
Non-organic hearing loss: new and confirmed findings.Eur Arch Otorhinolaryngol. 2010 Aug;267(8):1213-9. doi: 10.1007/s00405-010-1218-y. Epub 2010 Mar 4. Eur Arch Otorhinolaryngol. 2010. PMID: 20204391
-
Influence of Compression Thresholds and Maximum Power Output on Speech Understanding with Bone-Anchored Hearing Systems.Biomed Res Int. 2021 Oct 22;2021:1518385. doi: 10.1155/2021/1518385. eCollection 2021. Biomed Res Int. 2021. PMID: 34722757 Free PMC article.
-
Speech intelligibility and its relation to auditory temporal processing in Czech and Swiss German subjects with and without tinnitus.Eur Arch Otorhinolaryngol. 2024 Mar;281(3):1589-1595. doi: 10.1007/s00405-023-08398-8. Epub 2024 Jan 4. Eur Arch Otorhinolaryngol. 2024. PMID: 38175264
-
Correlation between Speech Perception Outcomes after Cochlear Implantation and Postoperative Acoustic and Electric Hearing Thresholds.J Clin Med. 2021 Jan 17;10(2):324. doi: 10.3390/jcm10020324. J Clin Med. 2021. PMID: 33477270 Free PMC article.
-
Potentially Inadequate Real-Life Speech Levels by Healthcare Professionals during Communication with Older Inpatients.Int J Environ Res Public Health. 2023 Mar 3;20(5):4543. doi: 10.3390/ijerph20054543. Int J Environ Res Public Health. 2023. PMID: 36901552 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous