Case Reports
. 2005 Oct;101(4):1094-1097.
doi: 10.1213/01.ane.0000167637.36187.22.

Difficult ventilation with a double-lumen endotracheal tube: an unusual manufacturing defect

Affiliations
Case Reports

Difficult ventilation with a double-lumen endotracheal tube: an unusual manufacturing defect

Hung-Shu Chen et al. Anesth Analg. 2005 Oct.

Abstract

We report an unexpected difficult ventilation with a double-lumen endotracheal tube in a patient receiving left upper lobe lung mass resection. The manufacturing defect in both limbs of the Opti-Port Right Angle Double Swivel Connector of the double-lumen tube resulted in this problem. This defect is difficult to localize by the usual recommended methods. We discuss a modified algorithm for difficult ventilation with a double-lumen endotracheal tube.

Implications: Difficult ventilation occurred during general anesthesia as the result of a manufacturing defect in both limbs of the connector of the double-lumen endotracheal tube. The problem was resolved with a careful approach, and there were no serious consequences.

PubMed Disclaimer

Comment in

References

    1. Scherer R, Lawin P. Pathophysiology and clinical aspects of single lung ventilation. Anasth Intensivther Notfallmed 1984;19:168–74.
    1. Benumof JL. The position of a double-lumen tube should be routinely determined by fiberoptic bronchoscopy. J Cardiothorac Vasc Anesth 1993;7:513–4.
    1. Smith GB, Hirsh NP, Ehrenwerth J. Placement of double-lumen endobronchial tubes. Br J Anaesth 1986;58:1317–20.
    1. Campbell C, Viswanathan S, Riopelle JM, Naraghi M. Manufacturing defect in a double-lumen tube. Anesth Analg 1991;73:825–6.
    1. Utting JE, Gray TC, Shelley FC. Human misadventure in anaesthesia. Can Anaesth Soc J 1979;26:472–8.

Publication types

LinkOut - more resources