Evaluation of a preprimed microporous hollow-fiber membrane for rapid response neonatal extracorporeal membrane oxygenation
- PMID: 16231623
- DOI: 10.1191/0267659105pf819oa
Evaluation of a preprimed microporous hollow-fiber membrane for rapid response neonatal extracorporeal membrane oxygenation
Abstract
Delays in initiating extracorporeal membrane oxygenation (ECMO) in the critically ill pediatric patient may lead to adverse outcomes. Maintaining a primed ECMO circuit can considerably reduce the initiation time. The predominant concerns precluding this practice are a decrease in oxygenator efficiency due to the saturation of microporous hollow fibers and compromised sterility when the oxygenator has been primed for 30 days. For institutions using a hollow-fiber oxygenator for ECMO, there are no data reporting pre-primed hollow-fiber oxygenator viability. This study reports the efficiency of oxygen transfer and the sterility of the Carmeda Minimax Plus (Medtronic, Inc, Minneapolis, MN) oxygenator after being crystalloid primed for 30 days. A total of 10 Minimax Plus oxygenators were tested for oxygen transfer in a laboratory setting utilizing fresh whole bovine blood. The control group (n =5) were tested immediately after priming. The test group (n =5) were oxygenators primed for 30 days with crystalloid solution and left stagnant until tested. Prior to testing, all oxygenators were circulated for 5 min and samples drawn to test for circuit sterility. Venous inlet saturations were manipulated to achieve three levels of testing: venous saturation (SvO2) of 55% for an oxygen challenge, SvO2 of 65% to comply with AAMI standards, and SvO2 of 75% to assess oxygen transfer rates and peak PaO2 achievement. Blood flow for all tests was maintained at 2 L/min with 1:1 blood to gas flow ratio and 100% FiO2. Samples were drawn pre- and postoxygenator at 1- and 6-hour time intervals to compute actual oxygen transfer values. All cultures from the test group priming solution produced no microbial growth after 30 days of stagnant prime. Average oxygen transfer values (ml/O2/min) for the control group after 1 hour of continuous use were 130.1+/-15.5 (@ 55% SvO2), 113.7+/-10.4 (@ 65% SvO2), 97.7+/-8.9 (@ 75% SvO2). After 6 hours, the average transfer values increased to 134.2+/-13.2 (@ 55% SvO2), 118.76+/-6.6 (@ 65% SvO2) and 98.9+/-8.3 (@ 75% SvO2). The average oxygen transfer values after 1 hour for oxygenators primed for 30 days were 114.9+/-10.0 (@ 55% SvO2), 112.4+/-8.2 (@ 65% SvO2) and 89.6+/-16.0 (@ 75% SvO2). After 6 hours of use, the average transfer values all decreased to 111.4+/-2.1 (@ 55% SvO2, p <0.05 versus control), 104.0+/-5.6 (@ 65% SvO2, p <0.05 versus control) and 88.4+/-3.2 (@ 75% SvO2, p <0.05 versus control). In conclusion, there was a decrease in the average oxygen transfer values for the test group after 6 hours versus the control. The modest loss of oxygen transfer ability observed can be considered acceptable due to the amount of surface area of the Minimax Plus oxygenator when used on a neonate, making it feasible to adopt the practice of prepriming the Minimax oxygenator for neonatal ECMO.
Similar articles
-
A wet-primed extracorporeal membrane oxygenation circuit with hollow-fiber membrane oxygenator maintains adequate function for use during cardiopulmonary resuscitation after 2 weeks on standby.Crit Care Med. 2005 Jul;33(7):1572-6. doi: 10.1097/01.ccm.0000168598.40541.22. Crit Care Med. 2005. PMID: 16003064
-
The effects of preprimed oxygenators on gas transfer efficiency.J Extra Corpor Technol. 2003 Jun;35(2):121-6. J Extra Corpor Technol. 2003. PMID: 12939020
-
Performance of a MedArray silicone hollow fiber oxygenator.ASAIO J. 2009 Jul-Aug;55(4):382-7. doi: 10.1097/MAT.0b013e3181a44263. ASAIO J. 2009. PMID: 19381081
-
Development of the oxygenator: past, present, and future.J Artif Organs. 2004;7(3):111-20. doi: 10.1007/s10047-004-0268-6. J Artif Organs. 2004. PMID: 15558331 Review.
-
Evolutions of extracorporeal membrane oxygenator (ECMO): perspectives for advanced hollow fiber membrane.J Artif Organs. 2024 Mar;27(1):1-6. doi: 10.1007/s10047-023-01389-w. Epub 2023 Mar 14. J Artif Organs. 2024. PMID: 36914927 Free PMC article. Review.
Cited by
-
Extracorporeal membrane oxygenation circuitry.Pediatr Crit Care Med. 2013 Jun;14(5 Suppl 1):S7-12. doi: 10.1097/PCC.0b013e318292dd10. Pediatr Crit Care Med. 2013. PMID: 23735989 Free PMC article.
-
Implications of pediatric extracorporeal cardiopulmonary resuscitation simulation for intensive care team confidence and coordination: A pilot study.Perfusion. 2024 Nov;39(8):1570-1576. doi: 10.1177/02676591231202679. Epub 2023 Sep 13. Perfusion. 2024. PMID: 37702710 Free PMC article.
-
Retrospective analysis comparing hollow fiber and silicone membrane oxygenators for neonates on ECMO.J Extra Corpor Technol. 2007 Jun;39(2):71-4. J Extra Corpor Technol. 2007. PMID: 17672186 Free PMC article.
-
Outcomes before and after implementation of a pediatric rapid-response extracorporeal membrane oxygenation program.Ann Thorac Surg. 2013 Jun;95(6):2140-6; discussion 2146-7. doi: 10.1016/j.athoracsur.2013.01.050. Epub 2013 Mar 15. Ann Thorac Surg. 2013. PMID: 23506632 Free PMC article.
-
Sterility Duration of Preprimed Extracorporeal Membrane Oxygenation Circuits.J Pediatr Pharmacol Ther. 2018 Jul-Aug;23(4):311-314. doi: 10.5863/1551-6776-23.4.311. J Pediatr Pharmacol Ther. 2018. PMID: 30181722 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources