Radiation dose reduction without degradation of low-contrast detectability at abdominal multisection CT with a low-tube voltage technique: phantom study
- PMID: 16237139
- DOI: 10.1148/radiol.2373041643
Radiation dose reduction without degradation of low-contrast detectability at abdominal multisection CT with a low-tube voltage technique: phantom study
Abstract
Purpose: To reduce radiation dose from abdominal computed tomography (CT) without degradation of low-contrast detectability by using a technique with low tube voltage (90 kV).
Materials and methods: The institutional review board approved the participation of the radiologists in the observer performance test, and informed consent was obtained from all participating radiologists. A phantom for measurement of the radiation dose and a phantom containing low-contrast objects were scanned with a 16-detector row CT scanner at 120 kV and 90 kV. For determination of the radiation dose at both 90 kV and 120 kV, the tube current-time product settings were 100-560 mAs, and the doses at the center and periphery of the phantom were measured. To assess low-contrast detectability, we used a 300-mAs setting at 120 kV and 250-560-mAs settings at 90 kV. Five observers participated in the receiver operating characteristic analysis. Area under the receiver operating characteristic curve (A(z)) values were calculated in each observer. A(z) values obtained with each of the scanning techniques were recorded, and differences were examined for significance by using the Dunnet method.
Results: The mean A(z) value was 0.951 at 120 kV and 300 mAs. A(z) values were 0.927-0.973 at 90 kV and 450-560 mAs, and the differences between those values and values obtained at 120 kV and 300 mAs were not significant (P = .937-.952). A value of 100% was assigned to the radiation dose delivered to the center of the phantom at 120 kV and 300 mAs. The relative dose delivered at 90 kV ranged from 65% at 450 mAs to 79% at 560 mAs.
Conclusion: A reduction from 120 kV to 90 kV led to as much as a 35% reduction in the radiation dose, without sacrifice of low-contrast detectability, at CT.
RSNA, 2005
Comment in
-
Automatic exposure control in CT: are we done yet?Radiology. 2005 Dec;237(3):755-6. doi: 10.1148/radiol.2373051151. Radiology. 2005. PMID: 16304094 No abstract available.
Similar articles
-
Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise.Radiology. 2005 Dec;237(3):945-51. doi: 10.1148/radiol.2373041655. Epub 2005 Oct 19. Radiology. 2005. PMID: 16237140 Clinical Trial.
-
Effect of low tube voltage on image quality, radiation dose, and low-contrast detectability at abdominal multidetector CT: phantom study.J Biomed Biotechnol. 2012;2012:130169. doi: 10.1155/2012/130169. Epub 2012 Apr 23. J Biomed Biotechnol. 2012. PMID: 22619490 Free PMC article.
-
Effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral pediatric and adult CT angiography: a phantom study.Invest Radiol. 2013 Apr;48(4):192-9. doi: 10.1097/RLI.0b013e31827efc17. Invest Radiol. 2013. PMID: 23344518
-
Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape.Radiology. 2004 Nov;233(2):515-22. doi: 10.1148/radiol.2332032107. Epub 2004 Sep 9. Radiology. 2004. PMID: 15358847
-
Reduction in radiation and contrast medium dose via optimization of low-kilovoltage CT protocols using a hybrid iterative reconstruction algorithm at 256-slice body CT: phantom study and clinical correlation.Clin Radiol. 2013 Mar;68(3):e128-35. doi: 10.1016/j.crad.2012.10.014. Epub 2012 Dec 11. Clin Radiol. 2013. PMID: 23245269
Cited by
-
Personalization of injection protocols to the individual patient's blood volume and automated tube voltage selection (ATVS) in coronary CTA.PLoS One. 2018 Sep 26;13(9):e0203682. doi: 10.1371/journal.pone.0203682. eCollection 2018. PLoS One. 2018. PMID: 30256835 Free PMC article.
-
CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction.BMC Med Imaging. 2013 Aug 9;13:27. doi: 10.1186/1471-2342-13-27. BMC Med Imaging. 2013. PMID: 23927627 Free PMC article.
-
Radiation dose reduction for CT lung cancer screening using ASIR and MBIR: a phantom study.J Appl Clin Med Phys. 2014 Mar 6;15(2):4515. doi: 10.1120/jacmp.v15i2.4515. J Appl Clin Med Phys. 2014. PMID: 24710436 Free PMC article.
-
Performance improvements of virtual monoenergetic images in photon-counting detector CT compared with dual source dual-energy CT: Fourier-based assessment.Phys Eng Sci Med. 2025 Mar;48(1):143-153. doi: 10.1007/s13246-024-01499-6. Epub 2024 Dec 10. Phys Eng Sci Med. 2025. PMID: 39658761 Free PMC article.
-
Prospective multicenter study on personalized and optimized MDCT contrast protocols: results on liver enhancement.Eur Radiol. 2021 Nov;31(11):8236-8245. doi: 10.1007/s00330-021-07953-3. Epub 2021 Apr 29. Eur Radiol. 2021. PMID: 33914115
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical