Plethysmograph and interrupter resistance measurements in prematurely born young children
- PMID: 16239293
- PMCID: PMC2672703
- DOI: 10.1136/adc.2005.079608
Plethysmograph and interrupter resistance measurements in prematurely born young children
Abstract
Background: Airways obstruction in premature infants is often assessed by plethysmography, which requires sedation. The interrupter (Rint) technique does not require sedation, but has rarely been examined in children under 2 years of age.
Objective: To compare Rint results with plethysmographic measurements of airway resistance (Raw) in prematurely born, young children.
Design: Prospective study.
Setting: Infant and Paediatric Lung Function Laboratories.
Patients: Thirty children with a median gestational age of 25-29 weeks and median postnatal age of 13 months.
Interventions and main outcome measures: The infants were sedated, airway resistance was measured by total body plethysmography (Raw), and Rint measurements were made using a MicroRint device. Further Raw and Rint measurements were made after salbutamol administration if the children remained asleep.
Results: Baseline measurements of Raw and Rint were obtained from 30 and 26 respectively of the children. Mean baseline Rint values were higher than mean baseline Raw results (3.45 v 2.84 kPa/l/s, p = 0.006). Limits of agreement for the mean difference between Rint and Raw were -1.52 to 2.74 kPa/l/s. Ten infants received salbutamol, after which the mean Rint result was 3.6 kPa/l/s and mean Raw was 3.1 kPa/l/s (limits of agreement -0.28 to 1.44 kPa/l/s).
Conclusion: The poor agreement between Rint and Raw results suggests that Rint measurements cannot substitute for plethysmographic measurements in sedated prematurely born infants.
Conflict of interest statement
Competing interests: none declared
References
-
- Yuksel B, Greenough A. Airway resistance and lung volume before and after bronchodilator therapy in symptomatic preterm infants. Respir Med 199488281–286. - PubMed
-
- Thomas M, Rafferty G, Limb E.et al Pulmonary function at follow up of very preterm infants from the UK oscillation study. Am J Respir Crit Care Med 200469868–872. - PubMed
-
- Carter E R, Stecenko A A, Pollock B H.et al Evaluation of the interrupter technique for the use of assessing airway obstruction in children. Pediatr Pulmonol 199417211–217. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical