Three-dimensional coronary reconstruction from routine single-plane coronary angiograms: in vivo quantitative validation
- PMID: 16243736
- DOI: 10.1080/14628840500193398
Three-dimensional coronary reconstruction from routine single-plane coronary angiograms: in vivo quantitative validation
Abstract
Background: Current X-ray technology displays the complex 3-dimensional (3-D) geometry of the coronary arterial tree as 2-dimensional (2-D) images. To overcome this limitation, an algorithm was developed for the reconstruction of the 3-D pathway of the coronary arterial tree using routine single-plane 2-D angiographic imaging. This method provides information in real-time and is suitable for routine use in the cardiovascular catheterization laboratory.
Objectives: The purpose of this study was to evaluate the precision of this algorithm and to compare it with 2-D quantitative coronary angiography (QCA) system.
Methods: Thirty-eight angiographic images were acquired from 11 randomly selected patients with coronary artery disease undergoing diagnostic cardiac catheterization. The 2-D images were analyzed using QCA software. For the 3-D reconstruction, an algorithm integrating information from at least two single-plane angiographic images taken from different angles was formulated.
Results: 3-D acquisition was feasible in all patients and in all selected angiographic frames. Comparison between pairs of values yielded greater precision of the 3-D than the 2-D measurements of the minimal lesion diameter (P<0.005), minimal lesion area (P<0.05) and lesion length (P<0.01).
Conclusions: The study validates the 3-D reconstruction algorithm, which may provide new insights into vessel morphology in 3-D space. This method is a promising clinical tool, making it possible for cardiologists to appreciate the complex curvilinear structure of the coronary arterial tree and to quantify atherosclerotic lesions more precisely.
Publication types
MeSH terms
LinkOut - more resources
Medical