Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an implantable loop recorder (ILR)

Affiliations
Clinical Trial

Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an implantable loop recorder (ILR)

J Rod Gimbel et al. Ann Noninvasive Electrocardiol. 2005 Oct.

Abstract

Background: Patients with implantable devices are generally not permitted to undergo magnetic resonance imaging (MRI) because of potentially deleterious interactions. Little has been reported regarding the safety and effects of MRI scanning of patients with implantable loop recorders (ILRs). We evaluated the safety of scanning patients with ILRs and the output of the ILR after undergoing MRI.

Methods: Ten patients underwent 11 MRI scanning events. All patients had Reveal Plus (Medtronic, Minneapolis, MN) ILRs. Seven cranial, two lumbar-spine, one shoulder, and one knee MRI were performed. All of the MRIs were performed with the understanding that the patient had an ILR. In each patient, the ILR was cleared moments before the scan and the integrity of the signal and time date stamp were verified. The devices were reinterrogated immediately after MRI in 10 patients and two days post MR scanning in one patient. Each patient was questioned post MRI regarding any symptoms experienced during the scan.

Results: Both tachy and bradyarrhythmias appeared as artifacts as a result of ILR exposure to MRI. Post MRI, none of the ILRs showed diminished signal integrity, altered programmed parameters, diminished battery status, inability to communicate or be reprogrammed. No sensations of tugging or warmth at the implant site were noted.

Conclusion: MRI was performed in ILR patients without harm to the patient or permanent damage to the ILR. MRI scanning of the Reveal appears safe. Artifact mimicking an arrhythmia was common, however, and must be excluded in any ILR patient undergoing MRI to avoid mistakenly attributing a syncopal episode, or palpitations to the artifacts produced from MRI exposure.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Artifact mimicking narrow complex tachycardia. Marked signal attenuation mimicking complete heart block. Both occurred during cranial MR scanning.
Figure 2
Figure 2
Prolonged episode of asystole recorded by ILR during lumbar spine MRI.

Similar articles

Cited by

References

    1. Krahn AD, Klein GJ, Skanes AC, et al Insertable loop recorder use for detection of intermittent arrhythmias. (Review) Pacing Clin Electrophysiol 2004;27(5):657–664. - PubMed
    1. Constantine G, Shan K, Flamm SD, et al Role of MRI in clinical cardiology. (Review) Lancet 2004;363(9427):2162–2171. - PubMed
    1. Shellock FG, Tkach JA, Ruggieri PM, et al Cardiac pacemakers, ICDs, and loop recorder: Evaluation of translational attraction using conventional (“long‐bore”) and “short‐bore” 1.5‐ and 3.0‐Tesla MR systems. J Cardiovasc Magn Reson 2003;5(2):387–397. - PubMed
    1. De Cock CC, Spruijt HJ, Van Campen LM, et al Electromagnetic interference of an implantable loop recorder by commonly encountered electronic devices. Pacing Clin Electrophysiol 2000;23(10 Pt 1):1516–1518. - PubMed
    1. Gimbel JR, Wilkoff B. Artefact mimicking tachycardia during magnetic resonance imaging in a patient with an implantable loop recorder. Heart 2003;89(3):e10. - PMC - PubMed

Publication types

MeSH terms