. 2005 Nov;24(11):1428-40.
doi: 10.1109/TMI.2005.856731.

A system for real-time XMR guided cardiovascular intervention

Affiliations

A system for real-time XMR guided cardiovascular intervention

Kawal S Rhode et al. IEEE Trans Med Imaging. 2005 Nov.

Abstract

The hybrid magnetic resonance (MR)/X-ray suite (XMR) is a recently introduced imaging solution that provides new possibilities for guidance of cardiovascular catheterization procedures. We have previously described and validated a technique based on optical tracking to register MR and X-ray images obtained from the sliding table XMR configuration. The aim of our recent work was to extend our technique by providing an improved calibration stage, real-time guidance during cardiovascular catheterization procedures, and further off-line analysis for mapping cardiac electrical data to patient anatomy. Specially designed optical trackers and a dedicated calibration object have resulted in a single calibration step that can be efficiently checked and updated before each procedure. An X-ray distortion model has been implemented that allows for distortion correction for arbitrary c-arm orientations. During procedures, the guidance system provides a real-time combined MR/X-ray image display consisting of live X-ray images with registered recently acquired MR derived anatomy. It is also possible to reconstruct the location of catheters seen during X-ray imaging in the MR derived patient anatomy. We have applied our registration technique to 13 cardiovascular catheterization procedures. Our system has been used for the real-time guidance of ten radiofrequency ablations and one aortic stent implantation. We demonstrate the real-time guidance using two exemplar cases. In a further two cases we show how off-line analysis of registered image data, acquired during electrophysiology study procedures, has been used to map cardiac electrical measurements to patient anatomy for two different types of mapping catheters. The cardiologists that have used the guidance system suggest that real-time XMR guidance could have substantial value in difficult interventional and electrophysiological procedures, potentially reducing procedure time and delivered radiation dose. Also, the ability to map measured electrical data to patient specific anatomy provides improved visualization and a path to investigation of cardiac electromechanical models.

PubMed Disclaimer

Similar articles

Cited by

Publication types

MeSH terms