Comparison of a needle-type and a microdialysis continuous glucose monitor in type 1 diabetic patients
- PMID: 16306547
- DOI: 10.2337/diacare.28.12.2871
Comparison of a needle-type and a microdialysis continuous glucose monitor in type 1 diabetic patients
Abstract
Objective: We examined the reliability of two continuous glucose sensors in type 1 diabetic patients at night and during rapid glucose excursions and verified the hypothesized nocturnal hypoglycemic drift of the needle-type sensor (CGMSgold) and delay of the microdialysis sensor (GlucoDay).
Research design and methods: Blood was sampled overnight twice per hour in 13 patients. Rapid-acting insulin was given subcutaneously 30 min after breakfast. Sampling once per minute started 45 min after breakfast and 75 min after insulin injection for 30 min, with the aim of determining peak and nadir glucose values. Mean absolute differences (MADs) between sensor and blood glucose values were calculated. Sensor curves were modeled for all patients using linear regression. Horizontal and vertical shifts of sensor curves from the blood glucose curves were assessed. A vertical shift indicates sensor drift and a horizontal shift sensor delay.
Results: Drift was minimal in the needle-type and microdialysis sensors (-0.02 and -0.04 mmol/l). Mean +/- SD delay was 7.1 +/- 5.5 min for the microdialysis sensor (P < 0.001). MAD was 15.0% for the needle-type sensor and 13.6% for the microdialysis sensor (P = 0.013). After correction for the 7-min delay, the microdialysis MAD improved to 11.7% (P < 0.0001).
Conclusions: The microdialysis sensor was more accurate than the needle-type sensor, with or without correction for a 7-min delay. In contrast to the previous version, the current needle-type sensor did not exhibit nocturnal hypoglycemic drift. Continuous subcutaneous glucose sensors are valuable adjunctive tools for glucose trend analyses. However, considering the large MADs, individual sensor values should be interpreted with caution.
Similar articles
-
Relationship between interstitial and blood glucose in type 1 diabetes patients: delay and the push-pull phenomenon revisited.Diabetes Technol Ther. 2007 Apr;9(2):169-75. doi: 10.1089/dia.2006.0007. Diabetes Technol Ther. 2007. PMID: 17425443
-
Development of a miniaturized glucose monitoring system by combining a needle-type glucose sensor with microdialysis sampling method. Long-term subcutaneous tissue glucose monitoring in ambulatory diabetic patients.Diabetes Care. 1994 May;17(5):387-96. doi: 10.2337/diacare.17.5.387. Diabetes Care. 1994. PMID: 8062605
-
Do sensor glucose levels accurately predict plasma glucose concentrations during hypoglycemia and hyperinsulinemia?Diabetes Care. 2002 May;25(5):889-93. doi: 10.2337/diacare.25.5.889. Diabetes Care. 2002. PMID: 11978686
-
Nocturnal hypoglycaemia in type 1 diabetes--consequences and assessment.Diabetes Metab Res Rev. 2004 Nov-Dec;20 Suppl 2:S43-6. doi: 10.1002/dmrr.513. Diabetes Metab Res Rev. 2004. PMID: 15551300 Review.
-
Implantable continuous glucose sensors.Curr Diabetes Rev. 2008 Aug;4(3):169-74. doi: 10.2174/157339908785294406. Curr Diabetes Rev. 2008. PMID: 18690897 Review.
Cited by
-
Real-time glucose estimation algorithm for continuous glucose monitoring using autoregressive models.J Diabetes Sci Technol. 2010 Mar 1;4(2):391-403. doi: 10.1177/193229681000400221. J Diabetes Sci Technol. 2010. PMID: 20307401 Free PMC article.
-
Future of Automated Insulin Delivery Systems.Diabetes Technol Ther. 2017 Jun;19(S3):S67-S72. doi: 10.1089/dia.2017.0012. Diabetes Technol Ther. 2017. PMID: 28585877 Free PMC article. Review.
-
Hypoglycemia alarm enhancement using data fusion.J Diabetes Sci Technol. 2010 Jan 1;4(1):34-40. doi: 10.1177/193229681000400105. J Diabetes Sci Technol. 2010. PMID: 20167165 Free PMC article.
-
Contribution of an intrinsic lag of continuous glucose monitoring systems to differences in measured and actual glucose concentrations changing at variable rates in vitro.J Diabetes Sci Technol. 2010 Nov 1;4(6):1393-9. doi: 10.1177/193229681000400614. J Diabetes Sci Technol. 2010. PMID: 21129335 Free PMC article.
-
Continuous glucose monitoring: a review for behavioral researchers.Psychosom Med. 2012 May;74(4):356-65. doi: 10.1097/PSY.0b013e31825769ac. Psychosom Med. 2012. PMID: 22582333 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical