On the existence of low-energy photons (<150 keV) in the unflattened x-ray beam from an ordinary radiotherapeutic target in a medical linear accelerator
- PMID: 16306657
- DOI: 10.1088/0031-9155/50/23/015
On the existence of low-energy photons (<150 keV) in the unflattened x-ray beam from an ordinary radiotherapeutic target in a medical linear accelerator
Abstract
Low-energy photons (<150 keV) are essential for obtaining high quality x-ray radiographs. These photons are usually produced in the accelerator target, but are effectively absorbed by the flattening filter and, at least partially, by the target itself. Experimental proof is presented for the existence of low-energy photons in the unflattened x-ray beam produced by a 6 MeV electron beam normally incident on the thinner of the two existing ports of the all-Cu radiotherapeutic target of a Clinac 18 (Varian Associates) linear accelerator. A number of one-shot absorption measurements were carried out with 12 foils of Pb absorbers with thicknesses varying from 0.25 to 3 mm in steps of 0.25 mm arranged symmetrically around the central axis on a 7.2 cm radius circumference. A Kodak ECL film-screen-cassette combination was used as a detector in the absorption measurements, in which optical density was measured as a function of the thickness of the Pb absorbers. Two sets of absorption measurements were carried out: the first one with the Clinac 18 6 MV unflattened beam and the second one with the Clinac 600C 6 MV therapeutic counterpart beam. There is a striking difference between the two sets: the optical density versus Pb-absorber thickness curve shows a sharp increase in optical density at small absorber thicknesses in the case of the unflattened 6 MV x-ray beam as compared with a gently sloping dependence in the case of the 6 MV therapeutic beam. A semi-quantitative assessment of the low-energy photon contribution to the whole optical density/contrast is presented. A 0.85 mm thick Pb absorber intercepting the 6 MV unflattened x-ray beam eliminates almost totally the sharp peak in the optical density curve at small Pb-absorber thicknesses. This constitutes additional evidence for the existence of low-energy photons (<150 keV) in the unflattened 6 MV beam from the Cu therapeutic target.
Similar articles
-
Dosimetric characteristics of unflattened 6 MV photon beams of a clinical linear accelerator: a Monte Carlo study.Appl Radiat Isot. 2007 Sep;65(9):1029-36. doi: 10.1016/j.apradiso.2007.04.023. Epub 2007 Jun 9. Appl Radiat Isot. 2007. PMID: 17616465
-
A Monte Carlo study on neutron and electron contamination of an unflattened 18-MV photon beam.Appl Radiat Isot. 2009 Jan;67(1):55-60. doi: 10.1016/j.apradiso.2008.07.013. Epub 2008 Jul 25. Appl Radiat Isot. 2009. PMID: 18760613
-
Treatment-planning study of prostate cancer intensity-modulated radiotherapy with a Varian Clinac operated without a flattening filter.Int J Radiat Oncol Biol Phys. 2007 Aug 1;68(5):1567-71. doi: 10.1016/j.ijrobp.2007.04.025. Epub 2007 Jun 4. Int J Radiat Oncol Biol Phys. 2007. PMID: 17544596
-
Field flatness and symmetry of photon beams: review of the current recommendations.Technol Health Care. 2003;11(4):283-8. Technol Health Care. 2003. PMID: 14600339 Review.
-
Ultralow-energy excitations and prospects for spatially resolved spectroscopy.Microsc Microanal. 2004 Feb;10(1):28-33. doi: 10.1017/S1431927604040280. Microsc Microanal. 2004. PMID: 15306064 Review.
Cited by
-
Dosimetric properties of a beam quality-matched 6 MV unflattened photon beam.J Appl Clin Med Phys. 2012 Jul 5;13(4):3701. doi: 10.1120/jacmp.v13i4.3701. J Appl Clin Med Phys. 2012. PMID: 22766941 Free PMC article.
-
Unflattened photon beams from the standard flattening filter free accelerators for radiotherapy: Advantages, limitations and challenges.J Med Phys. 2011 Jul;36(3):123-5. doi: 10.4103/0971-6203.83464. J Med Phys. 2011. PMID: 21897556 Free PMC article. No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical