Randomized Controlled Trial
. 2005 May;10(3):165-71.
doi: 10.3109/10929080500229561.

A randomized controlled trial of human versus robotic and telerobotic access to the kidney as the first step in percutaneous nephrolithotomy

Affiliations
Free article
Randomized Controlled Trial

A randomized controlled trial of human versus robotic and telerobotic access to the kidney as the first step in percutaneous nephrolithotomy

Ben Challacombe et al. Comput Aided Surg. 2005 May.
Free article

Abstract

Objective: We present results from the first randomized controlled trial of human vs. telerobotic access to the kidney during percutaneous nephrolithotomy.

Methods: To compare (a) human with robotic percutaneous needle access and (b) local robotic with trans-Atlantic robotic percutaneous needle access, we used a validated kidney model into which a needle was inserted 304 times. Half the insertions were performed by a robotic arm and the other half by urological surgeons. Order was decided randomly except for a sub-group of 30 trans-Atlantic robotic procedures that were controlled by a team at Johns Hopkins, Baltimore, via four ISDN lines.

Results: All attempts were successful within three passes with a median time of 35 s for human attempts compared with a median of 57 s for robotic attempts. The robot was slower than the human to complete insertions (p < 0.001, Mann-Whitney U test), but was more accurate when compared with human operators as it made fewer attempts (88% robotic vs. 79% human first attempt success; p = 0.046, chi-squared test). Times for trans-Atlantic robotic needle insertion (median = 59 s) were comparable to times taken for local robotic needle insertion (median = 56 s) with no difference in accuracy.

Conclusion: Telerobotics is an accurate and feasible tool for future minimally invasive surgery.

PubMed Disclaimer

Publication types

LinkOut - more resources