Continuous dual oximetry in surgical critical care. Indications and limitations
- PMID: 1632703
- PMCID: PMC1242547
- DOI: 10.1097/00000658-199207000-00009
Continuous dual oximetry in surgical critical care. Indications and limitations
Abstract
Continuous dual oximetry combines pulse and venous oximetry to provide real-time information about oxygen utilization and pulmonary function. The authors undertook this study to examine the accuracy, utility, and limitations of the technique in surgical critical care. Twelve critically ill patients underwent placement of a modified pulmonary artery catheter and a pulse oximeter, both connected to an on-line computer. Paired blood gas and oximeter measurements were recorded every 4 to 6 hours, with a minimum of six sets per patient. Blood-gas-derived shunt fraction correlated well with oximeter-derived ventilation-perfusion index (r = 0.78, p less than 0.01). Further, the continuous oxygen extraction ratio (O2EI) and mixed venous oxygen saturation (SpvO2) correlated with the oxygen utilization coefficient (O2EI:r = 0.6, p less than 0.01; SpvO2, r = 0.76, p less than 0.01). Computer modeling of ventilation-perfusion index found limitations in accuracy that occur at high arterial oxygen saturations and when pulse oximetry errors are present. The authors conclude that (1) Continuous dual oximetry offers a significant advantage over routine blood gas analysis for monitoring cardiopulmonary parameters in critically ill patients because of its real-time display; (2) Oximeter-derived determinations parallel traditional parameters; and (3) The most significant technical limitations occur primarily at high arterial saturations and with pulse oximetry errors.
Similar articles
-
Continuous monitoring of gas exchange and oxygen use with dual oximetry.J Clin Anesth. 1988;1(1):3-8. doi: 10.1016/0952-8180(88)90003-7. J Clin Anesth. 1988. PMID: 3272741
-
Precision of noninvasive hemoglobin-level measurement by pulse co-oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds.Crit Care Med. 2012 Sep;40(9):2576-82. doi: 10.1097/CCM.0b013e318258fb4f. Crit Care Med. 2012. PMID: 22732278
-
Proof of concept non-invasive estimation of peripheral venous oxygen saturation.Biomed Eng Online. 2017 May 19;16(1):60. doi: 10.1186/s12938-017-0351-x. Biomed Eng Online. 2017. PMID: 28526082 Free PMC article.
-
[Arterial puncture or pulse oximetry?].Ugeskr Laeger. 1999 Feb 22;161(8):1100-2. Ugeskr Laeger. 1999. PMID: 10074849 Review. Danish.
-
Venous oximetry.Curr Opin Crit Care. 2006 Jun;12(3):263-8. doi: 10.1097/01.ccx.0000224872.09077.dc. Curr Opin Crit Care. 2006. PMID: 16672787 Review.
Cited by
-
Continuous central venous oxygen saturation (ScvO2) measurement using a fibre optic catheter in newborn infants.Arch Dis Child Fetal Neonatal Ed. 1996 May;74(3):F177-81. doi: 10.1136/fn.74.3.f177. Arch Dis Child Fetal Neonatal Ed. 1996. PMID: 8777680 Free PMC article.
-
Simultaneous in vivo comparison of two-versus three-wavelength mixed venous (Svo2) oximetry catheters.J Clin Monit. 1995 Sep;11(5):329-34. doi: 10.1007/BF01616992. J Clin Monit. 1995. PMID: 7595690
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical