Randomized Controlled Trial
. 2006 Jan;13(1):1-6.
doi: 10.1197/j.aem.2005.07.036. Epub 2005 Dec 19.

Influence of electromagnetic fields on function of automated external defibrillators

Affiliations
Free article
Randomized Controlled Trial

Influence of electromagnetic fields on function of automated external defibrillators

Roman Fleischhackl et al. Acad Emerg Med. 2006 Jan.
Free article

Abstract

Objectives: In this study, the authors tested whether electromagnetic interference (EMI) is able to impair correct electrocardiogram analysis and produce false-positive shock advice from automated external defibrillators (AEDs) when the true rhythm is sinus.

Methods: Nineteen healthy subjects were used to test five AEDs available on the Austrian market in a prospective, open, and sequence-randomized study. The primary outcome variable was the absolute number of shocks advised in the presence of EMI. The secondary outcome was the number of impaired analyses caused by incorrectly detected patient movements or electrode failure.

Results: Of 760 tests run, 18 (2.37%) cases of false-positive results occurred, and two of five AEDs recommended shocks in the presence of sinus rhythm. Of 760 tests run, no electrode failures occurred. There were 27 occurrences (3.55%) of motion detected by an AED in the presence of strong electromagnetic fields.

Conclusions: AED models differ in their response to EMI; it may be useful to consider specific safety requirements for areas with such fields present. Working personnel and emergency medical services staff should be informed about potential risks and the possible need for patient evacuation before AEDs are attached and shock recommendations are followed.