The influence of flow rate on breathing circuit compliance and tidal volume delivered to patients in mechanical ventilation
- PMID: 16365508
- DOI: 10.1088/0967-3334/27/1/003
The influence of flow rate on breathing circuit compliance and tidal volume delivered to patients in mechanical ventilation
Abstract
Assessment of the gas volume that actually reaches the airways during mechanical ventilation appears to be a difficult task because of the presence of the breathing circuit. Most ventilators measure tidal volume at ventilator level making the determination of circuit compliance a critical factor in estimating the actual tidal volume. Tubing compliance can be measured in several ways and its value, being strongly dependent on the compressibility of the gas, may significantly differ depending on the measurement procedure. This paper addresses the dependence of the circuit compliance on the flow rate, and a theoretical hypothesis on the pneumatic behaviour of the breathing circuit is formulated and experimentally validated, with several tests conducted in vitro on an infant analogue. The dependence of the compliance on the inspiratory flow is experimentally assessed, and differences of about 20% on the measured value in the common flow range utilized in infant ventilation have been found, with consequent estimation errors of the volume delivered. Experimental tests show that the correct value of the tidal volume actually delivered to the patient can be reliably estimated from measurements performed at the ventilator level if the circuit compliance is determined with the same flow rates that will be utilized in mechanical ventilation.
Similar articles
-
Do new anesthesia ventilators deliver small tidal volumes accurately during volume-controlled ventilation?Anesth Analg. 2008 May;106(5):1392-400, table of contents. doi: 10.1213/ane.0b013e31816a68c6. Anesth Analg. 2008. PMID: 18420850
-
[A study of the parameters of the delivered tidal volume. Ventilation on a lung model using the CICERO anesthetic ventilator].Anaesthesist. 1992 Dec;41(12):785-9. Anaesthesist. 1992. PMID: 1489078 German.
-
Using conventional infant ventilators at unconventional rates.Pediatrics. 1984 Oct;74(4):487-92. Pediatrics. 1984. PMID: 6384912
-
[What are the changes in paediatric anaesthesia practice afforded by new anaesthetic ventilators?].Ann Fr Anesth Reanim. 2006 Apr;25(4):417-23. doi: 10.1016/j.annfar.2005.10.003. Epub 2005 Nov 28. Ann Fr Anesth Reanim. 2006. PMID: 16310339 Review. French.
-
Respiratory mechanics derived from signals in the ventilator circuit.Respir Care. 2005 Jan;50(1):55-65; discussion 65-7. Respir Care. 2005. PMID: 15636645 Review.
Cited by
-
Go with the flow-clinical importance of flow curves during mechanical ventilation: A narrative review.Can J Respir Ther. 2020 Jul 22;56:11-20. doi: 10.29390/cjrt-2020-002. eCollection 2020. Can J Respir Ther. 2020. PMID: 32844110 Free PMC article.
-
Laminar flowmeter for mechanical ventilator: Manufacturing challenge of Covid-19 pandemic.Flow Meas Instrum. 2021 Dec;82:102058. doi: 10.1016/j.flowmeasinst.2021.102058. Epub 2021 Oct 1. Flow Meas Instrum. 2021. PMID: 34611384 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources