The transmission of stress to grafted bone inside a titanium mesh cage used in anterior column reconstruction after total spondylectomy: a finite-element analysis
- PMID: 16371903
- DOI: 10.1097/01.brs.0000192281.53603.3f
The transmission of stress to grafted bone inside a titanium mesh cage used in anterior column reconstruction after total spondylectomy: a finite-element analysis
Abstract
Study design: A finite-element study of posterior alone or anterior/posterior combined instrumentation following total spondylectomy and replacement with a titanium mesh cage used as an anterior strut.
Objectives: To compare the effect of posterior instrumentation versus anterior/posterior instrumentation on transmission of the stress to grafted bone inside a titanium mesh cage following total spondylectomy.
Summary of background data: The most recent reconstruction techniques following total spondylectomy for malignant spinal tumor include a titanium mesh cage filled with autologous bone as an anterior strut. The need for additional anterior instrumentation with posterior pedicle screws and rods is controversial. Transmission of the mechanical stress to grafted bone inside a titanium mesh cage is important for fusion and remodeling. To our knowledge, there are no published reports comparing the load-sharing properties of the different reconstruction methods following total spondylectomy.
Methods: A 3-dimensional finite-element model of the reconstructed spine (T10-L4) following total spondylectomy at T12 was constructed. A Harms titanium mesh cage (DePuy Spine, Raynham, MA) was positioned as an anterior replacement, and 3 types of the reconstruction methods were compared: (1) multilevel posterior instrumentation (MPI) (i.e., posterior pedicle screws and rods at T10-L2 without anterior instrumentation); (2) MPI with anterior instrumentation (MPAI) (i.e., MPAI [Kaneda SR; DePuy Spine] at T11-L1); and (3) short posterior and anterior instrumentation (SPAI) (i.e., posterior pedicle screws and rods with anterior instrumentation at T11-L1). The mechanical energy stress distribution exerted inside the titanium mesh cage was evaluated and compared by finite-element analysis for the 3 different reconstruction methods. Simulated forces were applied to give axial compression, flexion, extension, and lateral bending.
Results: In flexion mode, the energy stress distribution in MPI was higher than 3.0 x 10 MPa in 73.0% of the total volume inside the titanium mesh cage, while 38.0% in MPAI, and 43.3% in SPAI. In axial compression and extension modes, there were no remarkable differences for each reconstruction method. In left-bending mode, there was little stress energy in the cancellous bone inside the titanium mesh cage in MPAI and SPAI.
Conclusions: This experiment shows that from the viewpoint of stress shielding, the reconstruction method, using additional anterior instrumentation with posterior pedicle screws (MPAI and SPAI), stress shields the cancellous bone inside the titanium mesh cage to a higher degree than does the system using posterior pedicle screw fixation alone (MPI). Thus, a reconstruction method with no anterior fixation should be better at allowing stress for remodeling of the bone graft inside the titanium mesh cage.
Similar articles
-
Healing of autologous bone in a titanium mesh cage used in anterior column reconstruction after total spondylectomy.Spine (Phila Pa 1976). 2002 Jul 1;27(13):E329-33. doi: 10.1097/00007632-200207010-00024. Spine (Phila Pa 1976). 2002. PMID: 12131755
-
The stability of reconstruction methods after thoracolumbar total spondylectomy. An in vitro investigation.Spine (Phila Pa 1976). 1999 Aug 15;24(16):1634-8. doi: 10.1097/00007632-199908150-00003. Spine (Phila Pa 1976). 1999. PMID: 10472096
-
Biomechanical analysis of anterior poly-methyl-methacrylate reconstruction following total spondylectomy for metastatic disease.Spine (Phila Pa 1976). 2004 Oct 1;29(19):2096-12. doi: 10.1097/01.brs.0000141186.01992.56. Spine (Phila Pa 1976). 2004. PMID: 15454698
-
Beneficial influence of titanium mesh cage on infection healing and spinal reconstruction in hematogenous septic spondylitis: a retrospective analysis of surgical outcome of twenty-five consecutive cases and review of literature.Spine (Phila Pa 1976). 2008 Oct 1;33(21):E759-67. doi: 10.1097/BRS.0b013e318187875e. Spine (Phila Pa 1976). 2008. PMID: 18827680 Review.
-
[Long-term Titanium mesh cage implant failure after total en bloc spondylectomy due to spinal chondrosarcoma: case report and literature review].Rev Fac Cien Med Univ Nac Cordoba. 2025 Mar 31;82(1):154-164. doi: 10.31053/1853.0605.v82.n1.44999. Rev Fac Cien Med Univ Nac Cordoba. 2025. PMID: 40163833 Free PMC article. Review. Spanish.
Cited by
-
FOUR-LEVEL EN BLOC VERTEBRECTOMY: A NOVEL TECHNIQUE AND LITERATURE REVIEW.Acta Ortop Bras. 2018;26(6):406-410. doi: 10.1590/1413-785220182606180916. Acta Ortop Bras. 2018. PMID: 30774516 Free PMC article.
-
Compression and contact area of anterior strut grafts in spinal instrumentation: a biomechanical study.BMC Musculoskelet Disord. 2013 Aug 26;14:254. doi: 10.1186/1471-2474-14-254. BMC Musculoskelet Disord. 2013. PMID: 23971712 Free PMC article.
-
Comparison of short-segment monoaxial and polyaxial pedicle screw fixation combined with intermediate screws in traumatic thoracolumbar fractures: a finite element study and clinical radiographic review.Clinics (Sao Paulo). 2017 Oct;72(10):609-617. doi: 10.6061/clinics/2017(10)04. Clinics (Sao Paulo). 2017. PMID: 29160423 Free PMC article.
-
Anatomy-related risk factors for the subsidence of titanium mesh cage in cervical reconstruction after one-level corpectomy.Int J Clin Exp Med. 2015 May 15;8(5):7405-11. eCollection 2015. Int J Clin Exp Med. 2015. PMID: 26221282 Free PMC article.
-
En bloc spondylectomy reconstructions in a biomechanical in-vitro study.Eur Spine J. 2008 May;17(5):715-25. doi: 10.1007/s00586-008-0588-y. Epub 2008 Jan 15. Eur Spine J. 2008. PMID: 18196295 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials