A technique of using gated-CT images to determine internal target volume (ITV) for fractionated stereotactic lung radiotherapy
- PMID: 16376444
- DOI: 10.1016/j.radonc.2005.11.012
A technique of using gated-CT images to determine internal target volume (ITV) for fractionated stereotactic lung radiotherapy
Abstract
Background and purpose: To develop and evaluate a technique and procedure of using gated-CT images in combination with PET image to determine the internal target volume (ITV), which could reduce the planning target volume (PTV) with adequate target coverage.
Patients and methods: A skin marker-based gating system connected to a regular single slice CT scanner was used for this study. A motion phantom with adjustable motion amplitude was used to evaluate the CT gating system. Specifically, objects of various sizes/shapes, considered as virtual tumors, were placed on the phantom to evaluate the number of phases of gated images required to determine the ITV while taking into account tumor size, shape and motion. A procedure of using gated-CT and PET images to define ITV for patients was developed and was tested in patients enrolled in an IRB approved protocol.
Results: The CT gating system was capable of removing motion artifacts for target motion as large as 3-cm when it was gated at optimal phases. A phantom study showed that two gated-CT scans at the end of expiration and the end of inspiration would be sufficient to determine the ITV for tumor motion less than 1-cm, and another mid-phase scan would be required for tumors with 2-cm motion, especially for small tumors. For patients, the ITV encompassing visible tumors in all sets of gated-CT and regular spiral CT images seemed to be consistent with the target volume determined from PET images. PTV expanded from the ITV with a setup uncertainty margin had less volume than PTVs from spiral CT images with a 10-mm generalized margin or an individualized margin determined at fluoroscopy.
Conclusions: A technique of determining the ITV using gated-CT images was developed and was clinically implemented successfully for fractionated stereotactic lung radiotherapy.
Similar articles
-
The use of gated and 4D CT imaging in planning for stereotactic body radiation therapy.Med Dosim. 2007 Summer;32(2):92-101. doi: 10.1016/j.meddos.2007.01.006. Med Dosim. 2007. PMID: 17472888
-
Dose calculations accounting for breathing motion in stereotactic lung radiotherapy based on 4D-CT and the internal target volume.Radiother Oncol. 2008 Jan;86(1):55-60. doi: 10.1016/j.radonc.2007.11.022. Epub 2007 Dec 20. Radiother Oncol. 2008. PMID: 18082905
-
Benefit of respiration-gated stereotactic radiotherapy for stage I lung cancer: an analysis of 4DCT datasets.Int J Radiat Oncol Biol Phys. 2005 Jun 1;62(2):554-60. doi: 10.1016/j.ijrobp.2005.01.032. Int J Radiat Oncol Biol Phys. 2005. PMID: 15890600
-
PET/CT (and CT) instrumentation, image reconstruction and data transfer for radiotherapy planning.Radiother Oncol. 2010 Sep;96(3):288-97. doi: 10.1016/j.radonc.2010.07.009. Epub 2010 Aug 13. Radiother Oncol. 2010. PMID: 20709416 Review.
-
[Multimodalities imaging for target volume definition in radiotherapy].Bull Cancer. 2006 Dec;93(12):1175-82. Bull Cancer. 2006. PMID: 17182373 Review. French.
Cited by
-
In-house auto cutoff sensor device for radiotherapy machine to monitor patient movements.J Appl Clin Med Phys. 2008 Jun 23;9(3):82-89. doi: 10.1120/jacmp.v9i3.2800. J Appl Clin Med Phys. 2008. PMID: 18716594 Free PMC article.
-
Measurement of time delay for a prospectively gated CT simulator.J Med Phys. 2010 Apr;35(2):123-7. doi: 10.4103/0971-6203.62196. J Med Phys. 2010. PMID: 20589123 Free PMC article.
-
Design and Evaluation of a MEMS Magnetic Field Sensor-Based Respiratory Monitoring and Training System for Radiotherapy.Sensors (Basel). 2018 Aug 21;18(9):2742. doi: 10.3390/s18092742. Sensors (Basel). 2018. PMID: 30134526 Free PMC article.
-
Efficient approach for determining four-dimensional computed tomography-based internal target volume in stereotactic radiotherapy of lung cancer.Radiat Oncol J. 2013 Dec;31(4):247-51. doi: 10.3857/roj.2013.31.4.247. Epub 2013 Dec 31. Radiat Oncol J. 2013. PMID: 24501714 Free PMC article.
-
A contrast-oriented algorithm for FDG-PET-based delineation of tumour volumes for the radiotherapy of lung cancer: derivation from phantom measurements and validation in patient data.Eur J Nucl Med Mol Imaging. 2008 Nov;35(11):1989-99. doi: 10.1007/s00259-008-0875-1. Epub 2008 Jul 26. Eur J Nucl Med Mol Imaging. 2008. PMID: 18661128
MeSH terms
LinkOut - more resources
Full Text Sources
Medical