Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry
- PMID: 16378009
- DOI: 10.1097/01.ijg.0000196655.85460.d6
Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry
Abstract
Purpose: The dynamic contour tonometer (DCT, Pascal tonometer, Swiss Microtechnology AG, Port, Switzerland) was recently introduced as a new method of intraocular pressure measurement, supposedly independent of corneal properties. In this study we analyzed the agreement and correlation of dynamic contour tonometry and Goldmann applanation tonometry (GAT) and investigated the influence of central corneal thickness (CCT) and corneal curvature. We also considered preferential patient groups for both methods.
Methods: In a prospective study of 100 eyes without glaucoma, intraocular pressure was measured using dynamic contour tonometry and Goldmann applanation tonometry, followed by measurements of central corneal thickness and corneal curvature.
Results: A clear correlation between dynamic contour tonometry and Goldmann applanation tonometry was found (r = 0.693; P < 0.001). Dynamic contour tonometry generally resulted in higher intraocular pressure measurements (median difference + 1.8 mm Hg, mean difference + 2.34 mm Hg). Unlike dynamic contour tonometry, Goldmann applanation tonometry was remarkably affected by central corneal thickness, but neither method was significantly influenced by corneal curvature. Bland-Altman graphs showed remarkable disagreement between dynamic contour tonometry and Goldmann applanation tonometry, which could be partially explained by the influence of central corneal thickness on Goldmann applanation tonometry. To obtain valid readings, dynamic contour tonometry required a more extensive selection of patients than Goldmann applanation tonometry.
Conclusions: Dynamic contour tonometry seems to be a reliable method for intraocular pressure measurement, which unlike Goldmann applanation tonometry is not influenced by central corneal thickness. In clinical practice, advantages from dynamic contour tonometry can be expected for cooperative patients, outpatients, and patients with sufficient bilateral ocular fixation, whereas Goldmann applanation tonometry measurements are more reliable in case of patients with inadequate cooperation, poor vision, or nystagmus.
Comment in
-
Dynamic contour tonometer versus goldmann applanation tonometer.J Glaucoma. 2006 Oct;15(5):471. doi: 10.1097/01.ijg.0000212266.34982.79. J Glaucoma. 2006. PMID: 16988614 No abstract available.
Similar articles
-
Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonometry.Ophthalmology. 2007 Jan;114(1):20-6. doi: 10.1016/j.ophtha.2006.06.047. Epub 2006 Oct 27. Ophthalmology. 2007. PMID: 17070592
-
Correlation between Central Corneal Thickness and Intraocular Pressure Measured by Goldmann Applanation Tonometry or Pascal Dynamic Contour Tonometry.Klin Monbl Augenheilkd. 2015 Apr;232(4):414-8. doi: 10.1055/s-0035-1545792. Epub 2015 Apr 22. Klin Monbl Augenheilkd. 2015. PMID: 25902089
-
Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study.Graefes Arch Clin Exp Ophthalmol. 2005 Aug;243(8):763-7. doi: 10.1007/s00417-005-1124-y. Epub 2005 Mar 9. Graefes Arch Clin Exp Ophthalmol. 2005. PMID: 15756572
-
New ways to measure intraocular pressure.Curr Opin Ophthalmol. 2008 Mar;19(2):122-6. doi: 10.1097/ICU.0b013e3282f391ae. Curr Opin Ophthalmol. 2008. PMID: 18301285 Review.
-
The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis.Indian J Ophthalmol. 2024 May 1;72(Suppl 3):S398-S403. doi: 10.4103/IJO.IJO_907_23. Epub 2023 Dec 15. Indian J Ophthalmol. 2024. PMID: 38099358 Free PMC article. Review.
Cited by
-
A comparison between Goldmann applanation tonometry and dynamic contour tonometry after photorefractive keratectomy.Graefes Arch Clin Exp Ophthalmol. 2013 Feb;251(2):603-8. doi: 10.1007/s00417-012-2142-1. Epub 2012 Sep 2. Graefes Arch Clin Exp Ophthalmol. 2013. PMID: 22940796
-
The role of light in measuring ocular biomechanics.Eye (Lond). 2016 Feb;30(2):234-40. doi: 10.1038/eye.2015.263. Epub 2016 Jan 15. Eye (Lond). 2016. PMID: 26768916 Free PMC article.
-
The effect of contact lens-induced corneal edema on Goldmann applanation tonometry and dynamic contour tonometry.Graefes Arch Clin Exp Ophthalmol. 2009 Mar;247(3):371-5. doi: 10.1007/s00417-008-0954-9. Epub 2008 Oct 9. Graefes Arch Clin Exp Ophthalmol. 2009. PMID: 18843499
-
Goldmann Tonometer Prism with an Optimized Error Correcting Applanation Surface.Transl Vis Sci Technol. 2016 Sep 9;5(5):4. doi: 10.1167/tvst.5.5.4. eCollection 2016 Sep. Transl Vis Sci Technol. 2016. PMID: 27642540 Free PMC article.
-
Evaluation of the new digital goldmann applanation tonometer for measuring intraocular pressure.J Ophthalmol. 2014;2014:461681. doi: 10.1155/2014/461681. Epub 2014 Jul 10. J Ophthalmol. 2014. PMID: 25126423 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources