Screening for duct-dependant congenital heart disease with pulse oximetry: a critical evaluation of strategies to maximize sensitivity

Affiliations

Screening for duct-dependant congenital heart disease with pulse oximetry: a critical evaluation of strategies to maximize sensitivity

Anne de Wahl Granelli et al. Acta Paediatr. 2005 Nov.

Abstract

Aim: To evaluate the feasibility of detecting duct-dependent congenital heart disease before hospital discharge by using pulse oximetry.

Design: Case-control study.

Setting: A supra-regional referral centre for paediatric cardiac surgery in Sweden.

Patients: 200 normal term newborns with echocardiographically normal hearts (median age 1.0 d) and 66 infants with critical congenital heart disease (CCHD; median age 3 d).

Methods: Pulse oximetry was performed in the right hand and one foot using a new-generation pulse oximeter (NGoxi) and a conventional-technology oximeter (CToxi).

Results: With the NGoxi, normal newborns showed a median postductal saturation of 99% (range 94-100%); intra-observer variability showed a mean difference of 0% (SD 1.3%), and inter-observer variability was 0% (SD 1.5%). The CToxi recorded a significantly greater proportion of postductal values below 95% (41% vs 1%) in the normal newborns compared with NGoxi (p<0.0001). The CCHD group showed a median postductal saturation of 90% (45-99%) with the NGoxi. Analysis of distributions suggested a screening cut-off of <95%; however, this still gave 7/66 false-negative patients, all with aortic arch obstruction. Best sensitivity was obtained by adding one further criterion: saturation of <95% in both hand and foot or a difference of >+/-3% between hand and foot. These combined criteria gave a sensitivity of 98.5%, specificity of 96.0%, positive predictive value of 89.0% and negative predictive value of 99.5%.

Conclusion: Systematic screening for CCHD with high accuracy requires a new-generation oximeter, and comparison of saturation values from the right hand and one foot substantially improves the detection of CCHD.