Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2006 Jan 3;47(1):175-81.
doi: 10.1016/j.jacc.2005.07.071. Epub 2005 Dec 1.

Hypersensitivity cases associated with drug-eluting coronary stents: a review of available cases from the Research on Adverse Drug Events and Reports (RADAR) project

Affiliations
Free article

Hypersensitivity cases associated with drug-eluting coronary stents: a review of available cases from the Research on Adverse Drug Events and Reports (RADAR) project

Jonathan R Nebeker et al. J Am Coll Cardiol. .
Free article

Abstract

Objectives: We undertook the review of all available cases of hypersensitivity reactions after placement of a drug-eluting stent (DES) and classified potential causes.

Background: Six months after the approval of the first DES, the Food and Drug Administration (FDA) reported 50 hypersensitivity reactions after stent placement but later concluded these were due to concomitantly prescribed medications such as clopidogrel. Nevertheless, the FDA continued to receive reports of hypersensitivity.

Methods: Reports available from April 2003 through December 2004 for hypersensitivity-like reactions associated with the sirolimus-eluting stent (CYPHER, Cordis Corp., Miami Lakes, Florida) and paclitaxel-eluting stent (TAXUS, Boston Scientific Corp., Natick, Massachusetts) were reviewed. Sources of reports included the FDA's adverse-device-event database, the published literature, and investigators from the Research on Adverse Drug/Device events And Reports (RADAR) project. Causality was assessed using standardized World Health Organization criteria.

Results: Of 5,783 reports identified for the DES in the FDA database, 262 unique events included hypersensitivity symptoms. Of these reports, 2 were certainly and 39 unlikely caused by clopidogrel and 1 was certainly, 9 probably, and 13 unlikely caused by the DES. From all sources, we identified 17 distinct cases that were probably or certainly caused by the stent, of which 9 had symptoms that lasted longer than four weeks. Four autopsies confirmed intrastent eosinophilic inflammation, thrombosis, and lack of intimal healing.

Conclusions: The FDA reports and autopsy findings suggest that DES may be a cause of systemic and intrastent hypersensitivity reactions that, in some cases, have been associated with late thrombosis and death.