Comparison of three methods of measuring corneal thickness and anterior chamber depth

Affiliations
Clinical Trial

Comparison of three methods of measuring corneal thickness and anterior chamber depth

Wolf Buehl et al. Am J Ophthalmol. 2006 Jan.

Abstract

Purpose: To compare three different methods of measuring corneal thickness (CT) and anterior chamber depth (ACD).

Design: Prospective clinical trial (Medical University of Vienna, Austria).

Methods: Central CT (CCT), CT at four peripheral points, and central ACD were measured in 88 eyes of 44 healthy subjects with the Pentacam (rotating Scheimpflug camera; Oculus, Wetzlar, Germany), Orbscan I (scanning-slit topography system; Orbtek Inc, Salt Lake City, Utah, USA), and AC-Master (partial coherence interferometry; Zeiss Meditec, Jena, Germany), and the results were compared.

Results: The upper (lower) limits of agreement for CCT measurements were 7.9 (-22.2) microm between AC-Master and Pentacam, 17.6 (-32.5) microm between AC-Master and Orbscan, and 25.2 (-25.9) microm between Pentacam and Orbscan. Correlation was high between all three methods (r = 0.94 to 0.97). The upper and lower limits of agreement for ACD were 0.174 (-0.251) mm between AC-Master and Pentacam, 0.406 (-0.004) mm between AC-Master and Orbscan, and 0.384 (0.095) mm between Pentacam and Orbscan. Correlation was high between the three methods (r = 0.96 between Orbscan and Pentacam; others 0.92). Correlation was lower for the CT measurements at the four peripheral points.

Conclusions: The CCT and ACD values obtained by Pentacam, Orbscan, and AC-Master measurements correlated well and showed few outliers. The two new systems (Pentacam, AC-Master) provide a reliable, easy-to-use, noncontact method of measuring CCT and ACD. Larger differences occurred only when measuring peripheral CT values, especially between AC-Master and the other two methods.