Efficacy of Cyto-Chex blood preservative for delayed manual CD4 testing using Dynal T4 Quant CD4 test among HIV-infected persons in Zambia
- PMID: 16394848
- DOI: 10.1097/01.qai.0000191282.77954.fc
Efficacy of Cyto-Chex blood preservative for delayed manual CD4 testing using Dynal T4 Quant CD4 test among HIV-infected persons in Zambia
Abstract
Background: Manual CD4 tests such as Dynal T4 Quant (Dynabeads, Dynal Biotech, Oslo, Norway) are less expensive alternatives to flow cytometry in resource-limited countries. Whereas blood preservatives have proven useful for stabilizing blood samples to allow delayed CD4 testing by flow cytometry, they have not been verified for manual tests. A method for preservation of blood prior to manual CD4 testing is needed for long-distance transport or sample batching.
Methods: Blood from HIV-positive Zambian military beneficiaries was mixed (1:1) with Cyto-Chex (Streck Laboratories, La Vista, NE) blood preservative, and the blood was stored at refrigerated, ambient, and incubator (37 degrees C) temperatures prior to Dynabeads CD4 testing at 0, 3, 6, and 9 days after collection. Baseline flow cytometry and Dynabeads testing without preservative were performed for comparison.
Results: Twenty-seven patient samples were analyzed. Dynabeads vs. flow cytometry had a correlation coefficient (r) of 0.84. There was excellent correlation (r = 0.96) between baseline Dynabeads testing and Cyto-Chex-preserved samples. Refrigerated samples showed strong correlation with baseline Dynabeads (r = 0.93-0.95) on days 3, 6, and 9 without decline in CD4 count (P = 0.73). Samples stored at ambient temperature yielded inferior results (r = 0.76-0.81), with a significant decline in CD4 count by day 3 (P < 0.001). The incubator arm had especially poor correlation (r = 0.30-0.49).
Conclusions: Addition of Cyto-Chex to peripheral blood (1:1) adequately preserves refrigerated blood samples for up to 9 days for subsequent testing with Dynabeads CD4 test. Cyto-Chex, however, cannot be recommended for delayed Dynabeads CD4 testing with storage at 37 degrees C or ambient temperatures in tropical areas similar to the site of this study.
Comment in
-
Cyto-Chex BCT, not Cyto-Chex, should be used for preservation of CD4 cell counts.J Acquir Immune Defic Syndr. 2006 Dec 1;43(4):503-4; author reply 504. doi: 10.1097/01.qai.0000242462.66194.ec. J Acquir Immune Defic Syndr. 2006. PMID: 17099318 No abstract available.
Similar articles
-
Evaluation of the blood stabilizers TransFix and Cyto-Chex BCT for low-cost CD4 T-cell methodologies.Viral Immunol. 2009 Oct;22(5):329-32. doi: 10.1089/vim.2009.0027. Viral Immunol. 2009. PMID: 19811090
-
Evaluation of Dynabeads and Cytospheres compared with flow cytometry to enumerate CD4+ T cells in HIV-infected Ugandans on antiretroviral therapy.J Acquir Immune Defic Syndr. 2008 Jul 1;48(3):297-303. doi: 10.1097/QAI.0b013e31817bbc3a. J Acquir Immune Defic Syndr. 2008. PMID: 18545154 Free PMC article.
-
Cyto-Chex BCT, not Cyto-Chex, should be used for preservation of CD4 cell counts.J Acquir Immune Defic Syndr. 2006 Dec 1;43(4):503-4; author reply 504. doi: 10.1097/01.qai.0000242462.66194.ec. J Acquir Immune Defic Syndr. 2006. PMID: 17099318 No abstract available.
-
Evaluation of a flow cytometry method for CD4 T cell enumeration based on volumetric primary CD4 gating using thermoresistant reagents.J Immunol Methods. 2011 Sep 30;372(1-2):7-13. doi: 10.1016/j.jim.2011.07.012. Epub 2011 Jul 29. J Immunol Methods. 2011. PMID: 21835181 Review.
-
Absolute and percent CD4+ T-cell enumeration by flow cytometry using capillary blood.J Immunol Methods. 2011 Sep 30;372(1-2):1-6. doi: 10.1016/j.jim.2011.07.008. Epub 2011 Jul 20. J Immunol Methods. 2011. PMID: 21787779 Review.
Cited by
-
Integration of semiconductor quantum dots into nano-bio-chip systems for enumeration of CD4+ T cell counts at the point-of-need.Lab Chip. 2008 Dec;8(12):2079-90. doi: 10.1039/b817116e. Epub 2008 Nov 5. Lab Chip. 2008. PMID: 19023471 Free PMC article.
-
Interlaboratory comparison of the TransFix®/EDTA Vacuum Blood Collection tube with the 5 mL Cyto-Chex® BCT.Cytometry B Clin Cytom. 2019 Nov;96(6):496-507. doi: 10.1002/cyto.b.21731. Epub 2018 Oct 24. Cytometry B Clin Cytom. 2019. PMID: 30353647 Free PMC article.
-
Multisite evaluation of point of care CD4 testing in Papua New Guinea.PLoS One. 2014 Nov 26;9(11):e112173. doi: 10.1371/journal.pone.0112173. eCollection 2014. PLoS One. 2014. PMID: 25426710 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials