Evaluation of a point-of-care direct spectrophotometric method for measurement of total serum bilirubin in term and near-term neonates
- PMID: 16407962
- DOI: 10.1038/sj.jp.7211436
Evaluation of a point-of-care direct spectrophotometric method for measurement of total serum bilirubin in term and near-term neonates
Abstract
Objective: To evaluate point-of-care (POC) measurement of total serum bilirubin (TSB) in the management of neonatal jaundice.
Study design: TSB was measured by a POC direct spectrophotometric bilirubin method (Unistat (U/TSB)) and a standard diazo clinical laboratory method (Olympus AU640E analyzer (diazo/TSB)). Agreement between U/TSB and diazo/TSB was assessed by correlation coefficient and Bland-Altman analysis. Transcutaneous bilirubin (TcB) was measured using JM-103 (JM).
Results: Correlation between U/TSB and diazo/TSB was 0.99 (n = 120). Maximum difference (U/TSB minus diazo/TSB) was -2.9 mg/dl, and 79% were +/-1 mg/dl; the average difference was -0.37+/-0.70 mg/dl and the average absolute difference was 0.60+/-0.52 mg/dl. Median time to determine U/TSB was 5 min. Correlation between U/TSB and JM was 0.92 (n = 113). Maximum difference (U/TSB minus JM) was 6.3 mg/dl, and 45% were +/-1 mg/dl; the average difference was 0.7+/-1.8 mg/dl and the average absolute difference was 1.4+/-1.2 mg/dl.
Conclusion: Measurement of TSB using Unistat provides excellent agreement with diazo/TSB and rapid turnaround time. This technique may provide reliable POC confirmation of TcB results that are above a screening cutoff value.
Similar articles
-
Evaluation of a transcutaneous jaundice meter following hospital discharge in term and near-term neonates.J Perinatol. 2005 Jul;25(7):486-90. doi: 10.1038/sj.jp.7211333. J Perinatol. 2005. PMID: 15908989 Clinical Trial.
-
Comparison of two transcutaneous bilirubinometers--Minolta AirShields Jaundice Meter JM103 and Spectrx Bilicheck--in Thai neonates.Southeast Asian J Trop Med Public Health. 2005 Nov;36(6):1533-7. Southeast Asian J Trop Med Public Health. 2005. PMID: 16610658
-
Noninvasive transcutaneous bilirubin as a screening test to identify the need for serum bilirubin assessment.J Med Assoc Thai. 2004 Oct;87(10):1193-8. J Med Assoc Thai. 2004. PMID: 15560697 Clinical Trial.
-
Point of care testing: transcutaneous bilirubinometry in neonates.Clin Biochem. 2009 Feb;42(3):143-9. doi: 10.1016/j.clinbiochem.2008.09.106. Epub 2008 Oct 2. Clin Biochem. 2009. PMID: 18929553 Review.
-
[Transcutaneous bilirubin measurement in neonates].An Esp Pediatr. 1993 Jan;38(1):33-7. An Esp Pediatr. 1993. PMID: 8439076 Review. Spanish.
Cited by
-
Correlation of microbilirubin with total serum bilirubin and transcutaneous bilirubin.PLoS One. 2025 Jun 11;20(6):e0324201. doi: 10.1371/journal.pone.0324201. eCollection 2025. PLoS One. 2025. PMID: 40498757 Free PMC article.
-
Point-of-care device to diagnose and monitor neonatal jaundice in low-resource settings.Proc Natl Acad Sci U S A. 2017 Dec 19;114(51):E10965-E10971. doi: 10.1073/pnas.1714020114. Epub 2017 Dec 4. Proc Natl Acad Sci U S A. 2017. PMID: 29203650 Free PMC article.
-
Modified Kramer's versus Kramer's Method for Clinical Assessment of Jaundice in Term and Near-Term Neonates.Indian J Pediatr. 2025 Apr;92(4):351-357. doi: 10.1007/s12098-023-05000-1. Epub 2024 Jan 12. Indian J Pediatr. 2025. PMID: 38214829
-
Development and Validation of a Smartphone Application for Neonatal Jaundice Screening.JAMA Netw Open. 2024 Dec 2;7(12):e2450260. doi: 10.1001/jamanetworkopen.2024.50260. JAMA Netw Open. 2024. PMID: 39661385 Free PMC article.
-
Feasibility study of texture-based machine learning approach for early detection of neonatal jaundice.Sci Rep. 2025 Feb 22;15(1):6481. doi: 10.1038/s41598-025-89528-6. Sci Rep. 2025. PMID: 39987322 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources