Determination of tetracycline, chlortetracycline and oxytetracycline by flow injection with inhibitory chemiluminescence detection using copper(II) as a probe ion
- PMID: 16416504
- DOI: 10.1002/bio.893
Determination of tetracycline, chlortetracycline and oxytetracycline by flow injection with inhibitory chemiluminescence detection using copper(II) as a probe ion
Abstract
This paper reports an indirect flow-injection (FI) method for the determination of the tetracycline drugs (TCs), tetracycline (TC), chlortetracycline (CTC) and oxytetracycline (OTC), using copper(II) as a probe ion. The method was based on the inhibition caused by these TCs to the copper(II)-catalysed chemiluminescence (CL) reaction between luminol and H(2)O(2). The CL reaction was induced on-line and injection of the sample produced negative peaks as a result of the copper(II) complexation or displacement by the analytes. The height of the peaks was proportional to the drug concentration in the sample. The choice of the catalyst ion, the concentration of luminol, H(2)O(2) and copper(II) are discussed. The linear range was 3.6 x 10(-8)-1.0 x 10(-5), 1.1 x 10(-7)-1.0 x 10(-5) and 1.9 x 10(-7)-1.0 x 10(-5) mol/L for TC, CTC and OTC, respectively. The detection limit was 5.0 x 10(-9) mol/L for TC, 1.0 x 10(-8) mol/L for CTC and 2.0 x 10(-8) mol/L for OTC (3sigma), respectively. The method was applied to the determination of TCs in pharmaceutical preparations and human urine with recoveries in the range 95-105%.
Similar articles
-
A sensitive inhibition chemiluminescence method for the determination of trace tannic acid using the reaction of luminol-hydrogen peroxide catalysed by tetrasulphonated manganese phthalocyanine.Luminescence. 2007 Jan-Feb;22(1):46-52. doi: 10.1002/bio.925. Luminescence. 2007. PMID: 16874849
-
The determination of glutamine with flow-injection chemiluminescence detection and mechanism study.Luminescence. 2010 Jan-Feb;25(1):50-4. doi: 10.1002/bio.1143. Luminescence. 2010. PMID: 19630088
-
The determination of hydrogen peroxide generated from cigarette smoke with an ultrasensitive and highly selective chemiluminescence method.Anal Chim Acta. 2007 Oct 3;601(1):95-100. doi: 10.1016/j.aca.2007.08.018. Epub 2007 Aug 19. Anal Chim Acta. 2007. PMID: 17904474
-
[Recent progress in flow-injection chemiluminescence for pharmaceutical analysis].Guang Pu Xue Yu Guang Pu Fen Xi. 2006 May;26(5):816-20. Guang Pu Xue Yu Guang Pu Fen Xi. 2006. PMID: 16883844 Review. Chinese.
-
Analytical applications of flow injection with chemiluminescence detection--a review.Luminescence. 2001 Jan-Feb;16(1):1-23. doi: 10.1002/bio.607. Luminescence. 2001. PMID: 11180653 Review.
Cited by
-
Construction of ratiometric fluorescence MIPs probe for selective detection of tetracycline based on passion fruit peel carbon dots and europium.Mikrochim Acta. 2021 Aug 16;188(9):297. doi: 10.1007/s00604-021-04929-4. Mikrochim Acta. 2021. PMID: 34401956
-
A chemiluminescence reaction consisting of manganese(IV), sodium sulfite, and sulfur- and nitrogen-doped carbon quantum dots, and its application for the determination of oxytetracycline.Mikrochim Acta. 2020 Feb 28;187(3):191. doi: 10.1007/s00604-020-4168-0. Mikrochim Acta. 2020. PMID: 32108911
-
Quantitative fluorescent detection of tetracycline in animal-derived foods using quantum dots.Appl Microbiol Biotechnol. 2024 Dec 14;108(1):535. doi: 10.1007/s00253-024-13253-9. Appl Microbiol Biotechnol. 2024. PMID: 39673609 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources