. 2006 Mar:444:184-92.
doi: 10.1097/01.blo.0000201148.06454.ef.

Computer-assisted total knee arthroplasty using patient-specific templating

Affiliations

Computer-assisted total knee arthroplasty using patient-specific templating

M A Hafez et al. Clin Orthop Relat Res. 2006 Mar.

Abstract

Current techniques used for total knee arthroplasty rely on conventional instrumentation that violates the intramedullary canals. Accuracy of the instrumentation is questionable, and assembly and disposal of the numerous pieces is time consuming. Navigation techniques are more accurate, but their broad application is limited by cost and complexity. We aimed to prove a new concept of computer-assisted preoperative planning to provide patient-specific templates that can replace conventional instruments. Computed tomography-based planning was used to design two virtual templates. Using rapid prototyping technology, virtual templates were transferred into physical templates (cutting blocks) with surfaces that matched the distal femur and proximal tibia. We performed 45 total knee arthroplasties on 16 cadaveric and 29 plastic knees, including a comparative trial against conventional instrumentations. All operations were performed using patient-specific templates with no conventional instrumentations, intramedullary perforation, tracking, or registration. The mean time for bone cutting was 9 minutes with a surgical assistant and 11 minutes without an assistant. Computer-assisted analyses of six random computed tomography scans showed mean errors for alignment and bone resection within 1.7 degrees and 0.8 mm (maximum, 2.3 degrees and 1.2 mm, respectively). Patient-specific templates are a practical alternative to conventional instrumentations, but additional clinical validation is required before clinical use.

PubMed Disclaimer

Similar articles

Cited by

Publication types