Mechanical loads modulate tidal volume and lung washout during high-frequency percussive ventilation
- PMID: 16448933
- DOI: 10.1016/j.resp.2005.02.015
Mechanical loads modulate tidal volume and lung washout during high-frequency percussive ventilation
Abstract
High-frequency percussive ventilation (HFPV) has been proved useful in patients with acute respiratory distress syndrome. However, its physiological mechanisms are still poorly understood. The aim of this work is to evaluate the effects of mechanical loading on the tidal volume and lung washout during HFPV. For this purpose a single-compartment mechanical lung simulator, which allows the combination of three elastic and four resistive loads (E and R, respectively), underwent HFPV with constant ventilator settings. With increasing E and decreasing R the tidal volume/cumulative oscillated gas volume ratio fell, while the duration of end-inspiratory plateau/inspiratory time increased. Indeed, an inverse linear relationship was found between these two ratios. Peak and mean pressure in the model decreased linearly with increasing pulsatile volume, the latter to a lesser extent. In conclusion, elastic or resistive loading modulates the mechanical characteristics of the HFPV device but in such a way that washout volume and time allowed for diffusive ventilation vary agonistically.
Similar articles
-
Effects of mechanical load on flow, volume and pressure delivered by high-frequency percussive ventilation.Respir Physiol Neurobiol. 2004 Aug 20;142(1):81-91. doi: 10.1016/j.resp.2004.04.005. Respir Physiol Neurobiol. 2004. PMID: 15351306
-
Measurement of pulsatile tidal volume, pressure amplitude, and gas flow during high-frequency percussive ventilation, with and without partial cuff deflation.Respir Care. 2007 Jan;52(1):45-9. Respir Care. 2007. PMID: 17194317
-
Measurement of changes in respiratory mechanics during partial liquid ventilation using jet pulses.Crit Care Med. 2003 May;31(5):1435-41. doi: 10.1097/01.CCM.0000063041.94690.76. Crit Care Med. 2003. PMID: 12771615
-
Lung-protective ventilation strategies in neonatology: what do we know--what do we need to know?Crit Care Med. 2007 Mar;35(3):925-31. doi: 10.1097/01.CCM.0000256724.70601.3A. Crit Care Med. 2007. PMID: 17255875 Review.
-
Is high-frequency ventilation more beneficial than low-tidal volume conventional ventilation?Respir Care Clin N Am. 2006 Sep;12(3):437-51. doi: 10.1016/j.rcc.2006.05.004. Respir Care Clin N Am. 2006. PMID: 16952803 Review.
Cited by
-
On some factors determining the pressure drop across tracheal tubes during high-frequency percussive ventilation: a flow-independent model.J Clin Monit Comput. 2021 Aug;35(4):885-890. doi: 10.1007/s10877-020-00548-1. Epub 2020 Jun 25. J Clin Monit Comput. 2021. PMID: 32588314 Free PMC article.
-
Comparison of flow and gas washout characteristics between pressure control and high-frequency percussive ventilation using a test lung.Physiol Meas. 2018 Mar 15;39(3):035001. doi: 10.1088/1361-6579/aaaaa2. Physiol Meas. 2018. PMID: 29369819 Free PMC article.
-
A narrative review of advanced ventilator modes in the pediatric intensive care unit.Transl Pediatr. 2021 Oct;10(10):2700-2719. doi: 10.21037/tp-20-332. Transl Pediatr. 2021. PMID: 34765495 Free PMC article. Review.
-
High frequency percussive ventilation increases alveolar recruitment in early acute respiratory distress syndrome: an experimental, physiological and CT scan study.Crit Care. 2018 Jan 11;22(1):3. doi: 10.1186/s13054-017-1924-6. Crit Care. 2018. PMID: 29325586 Free PMC article.
-
Gas distribution in a two-compartment model ventilated in high-frequency percussive and pressure-controlled modes.Intensive Care Med. 2010 Dec;36(12):2125-31. doi: 10.1007/s00134-010-1993-3. Epub 2010 Aug 6. Intensive Care Med. 2010. PMID: 20689922
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources