Three-dimensional tibiofemoral articular contact kinematics of a cruciate-retaining total knee arthroplasty
- PMID: 16452753
- DOI: 10.2106/JBJS.D.03028
Three-dimensional tibiofemoral articular contact kinematics of a cruciate-retaining total knee arthroplasty
Abstract
Background: Accurate knowledge of the location of tibiofemoral articular contact following total knee arthroplasty is important in order to understand polyethylene wear and the mechanisms of component failure. The present study was performed to determine the three-dimensional tibiofemoral articular contact patterns of a posterior cruciate ligament-retaining total knee replacement during in vivo weight-bearing flexion.
Methods: Nine osteoarthritic patients who were managed with a single design of a posterior cruciate ligament-retaining total knee implant were investigated with the use of an innovative dual orthogonal fluoroscopic imaging system. The position of the components during in vivo weight-bearing flexion was measured from full extension to maximum flexion in 15 degrees intervals. Tibiofemoral articular contact was determined by the overlap of the tibiofemoral articular surfaces. The centroid of the surface intersection was used to report the point of contact location. The average tibiofemoral contact points on both the medial and lateral tibial component surfaces were reported as a function of flexion.
Results: The average maximum weight-bearing flexion angle was 113.3 degrees +/- 13.1 degrees (range, 96 degrees to 138 degrees ). In the anteroposterior direction, the contact location was relatively constant in the medial compartment and moved posteriorly by 5.6 mm in the lateral compartment as the knee flexed from full extension to 90 degrees of flexion. The range of the contact location in the mediolateral direction was 3.7 mm in the medial compartment and 4.8 mm in the lateral compartment. For both compartments, posterior translation of the contact point was significant from 90 degrees to maximum flexion, but the contact point at maximum flexion was not observed to reach the posterior edge of the polyethylene tibial insert articular surface.
Conclusions: While the minimum anteroposterior translation of the contact point on the medial side might be interpreted as a medial pivot rotation during knee flexion, the contact point did move in the mediolateral direction with flexion. Beyond 90 degrees , both medial and lateral contact points were shown to move posteriorly but stopped before reaching the posterior edge of the polyethylene tibial insert articular surface. It seemed that the current component design did not allow the femoral condyle to roll off the polyethylene edge at high degrees of flexion because of the geometry at the posterior lip.
Similar articles
-
Mobile-bearing insert translational and rotational kinematics in a PCL-retaining total knee arthroplasty.Orthop Traumatol Surg Res. 2009 Jun;95(4):254-9. doi: 10.1016/j.otsr.2009.03.012. Epub 2009 May 12. Orthop Traumatol Surg Res. 2009. PMID: 19442597
-
Anterior cruciate ligament deficiency alters the in vivo motion of the tibiofemoral cartilage contact points in both the anteroposterior and mediolateral directions.J Bone Joint Surg Am. 2006 Aug;88(8):1826-34. doi: 10.2106/JBJS.E.00539. J Bone Joint Surg Am. 2006. PMID: 16882908
-
The effect of geometry of the tibial polyethylene insert on the tibiofemoral contact kinematics in Advance Medial Pivot total knee arthroplasty.J Orthop Sci. 2009 Nov;14(6):754-60. doi: 10.1007/s00776-009-1402-3. Epub 2009 Dec 8. J Orthop Sci. 2009. PMID: 19997823
-
Influence of component design on in vivo tibiofemoral contact patterns during kneeling after total knee arthroplasty: a systematic review and meta-analysis.Knee Surg Sports Traumatol Arthrosc. 2021 Feb;29(2):446-466. doi: 10.1007/s00167-020-05949-y. Epub 2020 Apr 3. Knee Surg Sports Traumatol Arthrosc. 2021. PMID: 32242268
-
In vivo kinematics of medial unicompartmental osteoarthritic knees during activities of daily living.Knee. 2014;21 Suppl 1:S10-4. doi: 10.1016/S0968-0160(14)50003-8. Knee. 2014. PMID: 25382361 Review.
Cited by
-
In vivo flexion and kinematics of the knee after TKA: comparison of a conventional and a high flexion cruciate-retaining TKA design.Knee Surg Sports Traumatol Arthrosc. 2009 Feb;17(2):150-6. doi: 10.1007/s00167-008-0637-4. Epub 2008 Oct 7. Knee Surg Sports Traumatol Arthrosc. 2009. PMID: 18839144
-
[A new total knee arthroplasty with physiologically shaped surfaces. Part 2: First clinical results].Unfallchirurg. 2009 Feb;112(2):176-84. doi: 10.1007/s00113-008-1551-4. Unfallchirurg. 2009. PMID: 19212737 German.
-
In vivo contact kinematics and contact forces of the knee after total knee arthroplasty during dynamic weight-bearing activities.J Biomech. 2008 Jul 19;41(10):2159-68. doi: 10.1016/j.jbiomech.2008.04.021. Epub 2008 Jun 5. J Biomech. 2008. PMID: 18538328 Free PMC article.
-
Patient function after a posterior stabilizing total knee arthroplasty: cam-post engagement and knee kinematics.Knee Surg Sports Traumatol Arthrosc. 2008 Mar;16(3):290-6. doi: 10.1007/s00167-007-0467-9. Epub 2008 Jan 9. Knee Surg Sports Traumatol Arthrosc. 2008. PMID: 18196219
-
In vivo knee kinematics during high flexion after a posterior-substituting total knee arthroplasty.Int Orthop. 2010 Apr;34(4):497-503. doi: 10.1007/s00264-009-0777-2. Epub 2009 Apr 22. Int Orthop. 2010. PMID: 19387643 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical