Comparison of standard culture methods, a shell vial assay, and a DNA probe for the detection of herpes simplex virus
- PMID: 1645373
- PMCID: PMC269839
- DOI: 10.1128/jcm.29.3.650-652.1991
Comparison of standard culture methods, a shell vial assay, and a DNA probe for the detection of herpes simplex virus
Abstract
A nonradioactive, biotinylated herpes simplex virus (HSV) DNA probe, a shell vial (rabbit kidney cell) culture assay enhanced by a direct fluorescent (HSV monoclonal)-antibody stain at 16 to 20 h postinoculation, and conventional tube cultures with confirmation via HSV-specific (polyclonal antibody) immunoperoxidase assay were compared for 199 specimens. The predictive values of the positive results were 54.5% for the probe, 95.9% for the shell vial assay, and 100% for the conventional culture methods, while the predictive values of the negative tests were 68.1, 84.0, and 98.4%, respectively. We conclude that the DNA probe (sensitivity, 24.5%; specificity, 88.3%) and the shell vial assay (sensitivity, 66.2%; specificity, 98.4%) cannot be substituted for conventional tube culture techniques (sensitivity, 97.1%; specificity, 100%) in the routine identification of HSV in our laboratory.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources