. 2006 Feb 17;832(1):17-23.
doi: 10.1016/j.jchromb.2005.12.034.

Sensitive and specific LC-MS/MS method for the simultaneous measurements of viramidine and ribavirin in human plasma

Affiliations

Sensitive and specific LC-MS/MS method for the simultaneous measurements of viramidine and ribavirin in human plasma

Yifei Liu et al. J Chromatogr B Analyt Technol Biomed Life Sci. .

Abstract

Viramidine is a prodrug of ribavirin. To facilitate pharmacokinetics studies of viramidine in man, a sensitive and specific LC-MS/MS method for the simultaneous analyses of viramidine and ribavirin in human plasma was developed and validated. The method involved the addition of [13C]viramidine and [13C]ribavirin as internal standards, protein precipitation with acetonitrile, HPLC separation, and quantification by MS/MS system using positive electrospray ionization in the multiple reaction monitoring mode (MRM). The precursor-->product ion transitions were monitored at 245-->113, 250-->113, 244-->112, and 249-->112 for ribavirin, [13C]ribavirin, viramidine, and [13C]viramidine, respectively. The calibration curves for viramidine and ribavirin were linear over a concentration range of 1-1000 ng/mL. For both viramidine and ribavirin, the lower limit of quantification (LLOQ) was 1 ng/mL. For viramidine, intra- and inter-day analyses of QC samples at 1, 5, 250, and 1000 ng/mL indicated good precision (%CV between 1.0 and 7.0%) and accuracy (%bias between -4.3 and 5.2%). For ribavirin, intra- and inter-day analyses of QC samples at 1, 5, 250, and 1000 ng/mL indicated similar precision (%CV between 0.8 and 8.3%) and accuracy (%bias between -5.8 and 9.4%). Both viramidine and ribavirin were stable in human plasma stored at room temperature for at least 3 h, 4 degrees C for at least 6 h, and for at least three freeze-thaw cycles. This accurate and highly specific assay provides a useful method for evaluating the pharmacokinetics of viramidine and ribavirin in man following administration of viramidine.