Comparative Study
. 2006 Feb;44(2):400-5.
doi: 10.1128/JCM.44.2.400-405.2006.

High analytical sensitivity and low rates of inhibition may contribute to detection of Chlamydia trachomatis in significantly more women by the APTIMA Combo 2 assay

Affiliations
Comparative Study

High analytical sensitivity and low rates of inhibition may contribute to detection of Chlamydia trachomatis in significantly more women by the APTIMA Combo 2 assay

Max Chernesky et al. J Clin Microbiol. 2006 Feb.

Abstract

The clinical sensitivity of nucleic acid amplification tests may be determined by analytical sensitivity and inhibitors in patient samples. We established endpoints for detection of propagated Chlamydia trachomatis L2 434, diluted according to swab and urine protocols for APTIMA Combo 2 (AC2), ProbeTec ET (PT), and Amplicor (AMP) assays. AC2 was 1,000-fold more sensitive than PT and 10-fold more sensitive than AMP on mock swab specimens. For urine, AC2 analytical sensitivity was 100-fold greater than those of the other assays. Spiking an aliquot of each clinical-trial sample from 298 women demonstrated inhibition rates in first-void urine (FVU), cervical swabs (CS), and vaginal swabs (VS) of 12.1%, 12.8%, and 10.4% for AMP; 27.2%, 2%, and 2%, for PT; and 0.3%, 1.7%, and 1.3% for AC2. Inhibition of our C. trachomatis spike and the PT or AMP amplification controls from the manufacturers showed less than 50% correlation. Using an infected-patient reference standard (a specimen positive in at least two tests or a single test positive in two of three samples) in AC2, the VS identified 68/69 (98.6%) infected women compared to CS (89.9%) or FVU (81.2%). Significantly fewer women were identified by PT (65.2%, 63.8%, and 66.7%) or AMP (65.2%, 59.4%, and 56.5%) with the three specimens. By individual specimen type, AC2 confirmed virtually all PT- and AMP-positive specimens, but rates of AC2 confirmation by AMP or PT ranged from 62.9 to 80.3%. The AC2 test identified significantly more women infected with C. trachomatis (P = 0.001). Vaginal swabs appear to be the specimen of choice for screening.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
RLU and optical densities (OD) of mock swab and urine specimens containing 10-fold dilutions of C. trachomatis (L2) tested by AC2, PT, and AMP.

Similar articles

Cited by

References

    1. Boyadzhyan, B., T. Yashina, J. H. Yatabe, M. Patnaik, and C. S. Hill. 2004. Comparison of the APTIMA CT and GC assays with the APTIMA combo 2 assay, the Abbott LCx assay, and direct fluorescent-antibody and culture assays for detection of Chlamydia trachomatis and Neisseria gonorrhoeae. J. Clin. Microbiol. 42:3089-3093. - PMC - PubMed
    1. Carder, C., A. J. Robinson, C. Broughton, J. M. Stephenson, and G. L. Ridgway. 1999. Evaluation of self-taken samples for the presence of genital Chlamydia trachomatis infection in women using the ligase chain reaction assay. Int. J. STD AIDS 10:776-779. - PubMed
    1. Chan, E. L., K. Brandt, K. Olienus, N. Antonishyn, and G. B. Horsman. 2000. Performance characteristics of the Becton Dickinson ProbeTec System for direct detection of Chlamydia trachomatis and Neisseria gonorrhoeae in male and female urine specimens in comparison with the Roche Cobas System. Arch. Pathol. Lab. Med. 124:1649-1652. - PubMed
    1. Chernesky, M., D. Jang, H. Lee, J. D. Burczak, H. Hu, J. Sellors, S. J. Tomazic-Allen, and J. B. Mahony. 1994. Diagnosis of Chlamydia trachomatis infections in men and women by testing first-void urine by ligase chain reaction. J. Clin. Microbiol. 32:2682-2685. - PMC - PubMed
    1. Chong, S., D. Jang, X. Song, J. Mahony, A. Petrich, P. Barriga, and M. Chernesky. 2003. Specimen processing and concentration of Chlamydia trachomatis added can influence false-negative rates in the LCx assay but not in the APTIMA Combo 2 assay when testing for inhibitors. J. Clin. Microbiol. 41:778-782. - PMC - PubMed

MeSH terms

Substances