Elimination of residual blood flow-related signal in 3D volume-selective TSE arterial wall imaging using velocity-sensitive phase reconstruction
- PMID: 16456824
- DOI: 10.1002/jmri.20513
Elimination of residual blood flow-related signal in 3D volume-selective TSE arterial wall imaging using velocity-sensitive phase reconstruction
Abstract
Purpose: To improve 3D volume-selective turbo spin-echo (TSE) arterial wall imaging by introducing velocity phase sensitivity to the sequence.
Materials and methods: Slow or recirculating blood may give incomplete nulling in 3D vessel wall imaging, hindering differentiation between arterial wall and lumen. Reconstructed phase images are used to reduce the residual blood signal with postprocessing. A volume-selective 3D TSE (1) sequence with increased velocity sensitivity perpendicular to the slab was developed. Sensitivity was introduced by modifying the gradient waveforms such that residual signal from slow or recirculating blood is distinguished from the wall by the phase shift introduced. This was tested on a pulsatile flow phantom and the carotid artery wall of six healthy volunteers and 17 patients. Images were acquired on a Siemens Magnetom Sonata 1.5T scanner. A quantitative comparison of the lumen/wall contrast was made using images acquired with and without modification.
Results: Velocity sensitivity produces significantly reduced residual signal by intravoxel dephasing. Phantom "lumen/tissue" contrast improved by 0.10 (P < 0.01), with a similar change in vivo (0.69-0.73, P < 0.01). Postprocessing removal of the blood signal using phase signal provided further improvement.
Conclusion: Introducing velocity sensitivity reduces unwanted, potentially misleading residual blood in 3D volume-selective vessel wall imaging.
(c) 2006 Wiley-Liss, Inc.
Similar articles
-
3D volume-selective turbo spin echo for carotid artery wall imaging with navigator detection of swallowing.J Magn Reson Imaging. 2005 Oct;22(4):583-8. doi: 10.1002/jmri.20424. J Magn Reson Imaging. 2005. PMID: 16161101
-
Visualization of hemodynamics in intracranial arteries using time-resolved three-dimensional phase-contrast MRI.J Magn Reson Imaging. 2007 Mar;25(3):473-8. doi: 10.1002/jmri.20828. J Magn Reson Imaging. 2007. PMID: 17279504
-
Volume-selective 3D turbo spin echo imaging for vascular wall imaging and distensibility measurement.J Magn Reson Imaging. 2003 May;17(5):572-80. doi: 10.1002/jmri.10294. J Magn Reson Imaging. 2003. PMID: 12720267
-
Phase-derivative analysis in MR angiography: reduced Venc dependency and improved vessel wall detection in laminar and disturbed flow.J Magn Reson Imaging. 1997 Mar-Apr;7(2):321-30. doi: 10.1002/jmri.1880070212. J Magn Reson Imaging. 1997. PMID: 9090586 Review.
-
Flow and motion.Magn Reson Imaging Clin N Am. 1999 Nov;7(4):699-715. Magn Reson Imaging Clin N Am. 1999. PMID: 10631674 Review.
Cited by
-
Multi-contrast high spatial resolution black blood inner volume three-dimensional fast spin echo MR imaging in peripheral vein bypass grafts.Int J Cardiovasc Imaging. 2010 Aug;26(6):683-91. doi: 10.1007/s10554-010-9621-4. Epub 2010 Mar 24. Int J Cardiovasc Imaging. 2010. PMID: 20333469 Free PMC article.
-
Bloody noise: The impact of blood-flow artifacts on registration.Hum Brain Mapp. 2023 Oct 1;44(14):4914-4926. doi: 10.1002/hbm.26426. Epub 2023 Jul 29. Hum Brain Mapp. 2023. PMID: 37516915 Free PMC article.
-
Accelerated 3D MERGE carotid imaging using compressed sensing with a hidden Markov tree model.J Magn Reson Imaging. 2012 Nov;36(5):1194-202. doi: 10.1002/jmri.23755. Epub 2012 Jul 23. J Magn Reson Imaging. 2012. PMID: 22826159 Free PMC article.
-
Dual stack black blood carotid artery CMR at 3T: application to wall thickness visualization.J Cardiovasc Magn Reson. 2009 Nov 10;11(1):45. doi: 10.1186/1532-429X-11-45. J Cardiovasc Magn Reson. 2009. PMID: 19903348 Free PMC article.
-
Improved blood suppression in three-dimensional (3D) fast spin-echo (FSE) vessel wall imaging using a combination of double inversion-recovery (DIR) and diffusion sensitizing gradient (DSG) preparations.J Magn Reson Imaging. 2010 Feb;31(2):398-405. doi: 10.1002/jmri.22042. J Magn Reson Imaging. 2010. PMID: 20099353 Free PMC article.