Verification of a rounded leaf-end MLC model used in a radiotherapy treatment planning system

Affiliations

Verification of a rounded leaf-end MLC model used in a radiotherapy treatment planning system

M J Williams et al. Phys Med Biol. .

Abstract

A new multileaf collimator (MLC) model has been incorporated into version 7.4 of the Pinnacle radiotherapy treatment planning system (Philips Radiation Oncology Systems, Milpitas, CA). The MLC model allows for rounded MLC leaf-ends and provides separate parameters for inter-leaf transmission, intra-leaf transmission and the tongue width of the MLC leaf. In this report we detail the method followed to commission the MLC model for a Varian 120-leaf Millennium MLC (Varian Medical Systems, Palo Alto, CA, USA) for both 6 and 10 MV photons, and test the validity of the model for an IMRT field. Dose profiles in water were measured for a range of square MLC field sizes and compared to the Pinnacle computed dose profiles; in addition, the dose distribution for a series of adjacent MLC fields was measured to observe the model's behaviour along match-lines. Based on these results intra-leaf transmissions of 1.5% for 6 MV and 1.8% for 10 MV, leaf-tip radius of 12.0 cm, an inter-leaf transmission of 0.5%, and a tongue width of 0.1 cm were chosen. Using these values to compute the planar dose distribution for a 6 MV IMRT field, the new version of Pinnacle displayed improved dosimetric agreement with the dose-to-water EPID image and ion chamber measurements when compared to the old version of Pinnacle, particularly along the MLC tongue edge and across match-lines. Discrepancies of up to 5% were observed between calculated and measured doses along match-lines for both 6 MV and 10 MV photons; however, the new MLC model did predict the presence of match-lines and was a significant improvement on the previous model.