Clinical Trial
. 2006 Feb;19(1):67-74.
doi: 10.1111/j.1540-8183.2006.00107.x.

Vascular complications with newer generations of angioseal vascular closure devices

Affiliations
Free article
Clinical Trial

Vascular complications with newer generations of angioseal vascular closure devices

Robert J Applegate et al. J Interv Cardiol. 2006 Feb.
Free article

Abstract

Objectives: To evaluate the safety and efficacy of newer generations of a single vascular closure device (VCD), Angioseal, compared to contemporaneous manual compression.

Background: Previous assessment of VCD safety and efficacy has been based primarily on outcomes associated with the original VCD. Few data exist comparing safety and efficacy of modifications of the original VCDs.

Methods: We evaluated in-hospital outcomes of patients undergoing diagnostic cath and percutaneous coronary intervention at a single center: 3,898 with manual compression and 3,898 with the Angioseal VCD. Three generations of the Angioseal device were studied: Generation 1, 7/97-4/00; Generation 2, 5/00-6/02; and Generation 3, 7/02-12/03. Propensity analysis was used to match patients within each generation for clinical and procedural covariates.

Results: Closure success was 98.5% for Generation 1, 98.6% for Generation 2, and 98.1% for Generation 3, P = NS. The odds ratio for any vascular complication comparing Angioseal to manual compression was 0.92 (0.42-2.01) for Generation 1; 0.83 (0.49-1.41) for Generation 2; and 0.75 (0.32-1.77) for Generation 3. Multivariate logistic regression identified closure success and Angioseal device Generation 3 versus Generation 1 as predictive of reduced vascular complications during the entire study period.

Conclusion: These data indicate that the newest generation of the Angioseal VCD is at least as effective as the original device. Moreover, use of the Angioseal VCD was associated with vascular complications similar to or lower than vascular complications following manual compression for each generation of Angioseal device analyzed.

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources