Conformability of balloon-expandable stents to the carotid siphon: an in vitro study
- PMID: 16484402
- PMCID: PMC8148779
Conformability of balloon-expandable stents to the carotid siphon: an in vitro study
Abstract
Background and purpose: Endovascular placement of coronary balloon-expandable stents in patients with recurrent cerebral ischemia has emerged as a treatment option for intracranial arterial occlusive disease. We have developed an in vitro model matching the tortuous curve of the carotid siphon that allows the assessment of apposition of stents to a curved vessel wall.
Methods: Six types of balloon-expandable coronary stents were implanted in a silicone model of the carotid siphon. Digital radiographs and 3D rotational angiograms were obtained. Stent morphology was evaluated and the degree of apposition between stent and wall of the model was measured on a digital workstation.
Results: All 6 stents showed lack of apposition between stent and the wall at the convexity of the anterior segment of the carotid siphon and the wall at the concavity at both extremities of the stent. In and around the curve, the modules of the stents did not expand completely to their nominal diameter and were distorted to an oval shape.
Conclusion: The tested coronary balloon-expandable stents did not completely conform to the vessel wall of the model of the carotid siphon and further development is needed to approach the goal of an "ideal intracranial stent."
Figures
References
-
- SSYLVIA Study Investigators. Stenting of Symptomatic Atherosclerotic Lesions in the Vertebral or Intracranial Arteries (SSYLVIA): study results. Stroke 2004;35:1388–92 - PubMed
-
- Mesnil de Rochemont R du, Turowski B, Buchkremer M, et al. Recurrent symptomatic high-grade intracranial stenoses: safety and efficacy of undersized stents: initial experience. Radiology 2004;231:45–49 - PubMed
-
- Rieu R, Barragan P, Garitey V, et al. Assessment of the trackability, flexibility, and conformability of coronary stents: a comparative analysis. Catheter Cardiovasc Interv 2003;59:496–503 - PubMed
-
- Kalmar G, Hubner F, Voelker W, et al. Radial force and wall apposition of balloon-expandable vascular stents in eccentric stenoses: an in vitro evaluation in a curved vessel model. J Vasc Interv Radiol 2002;13:499–508 - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources